Helio PM5500, PM5400 user manual Reverse Gas Flow Procedure, Operating Instructions

Page 10

Reverse Gas Flow Procedure

(CE Requirements ONLY)

1.Assure bleed flow valve is not engaged. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the HELIO2 Blender to ensure that there is no outlet flow.

2.Place the free end of the oxygen supply hose under water. While gradually increasing the heliox supply pressure from 30-75 psi

(2.07-5.17 bar) check for leakage past the oxygen inlet check valve.

3.Replace the Duckbill Check Valve in the oxygen inlet, if leakage is >100 ml/min. Reference HELIO2 Blender Service Manual (P/N 506124).

4.Repeat steps 1-3 to check for leakage past the heliox inlet check valve.

OPERATING INSTRUCTIONS

CAUTION

Inspect the HELIO2 Blender for visual damage before use, DO NOT USE if damaged.

1.Secure Blender to wall or pole mount bracket.

2.Connect heliox and oxygen supply lines from HELIO2 Blender to heliox cylinder and oxygen supply.

3.Connect oxygen flowmeter to Blender outlet.

4.Adjust the Oxygen Concentration Dial to the prescribed concentration. The balance of the concentration will be helium exiting the flowmeter.

NOTE: The Oxygen Concentration Dial does not rotate 360°. DO NOT force dial beyond the oxygen concentration endpoints as this will damage the Blender.

5.Confirm the flow of heliox mixture to the patient.

6.Actual heliox concentration to the patient may vary due to entrainment of room air via the patient interface device.

CAUTION

Refer to “Oxygen Flowmeter Conversion Chart” for corrected heliox flows.

An oxygen flowmeter should be used on the outlets of the HELIO2 Blender along with the corresponding flow conversion chart.

Actual flow from an oxygen flowmeter utilized to deliver heliox are higher than read on the flowmeter.

HELIO2

 

Blender

9

HELIUM-Oxygen

Image 10
Contents E R M a N U a L Model No. PM5400 Series shown PM5500 SeriesReceiving / Inspection ContentsIntended Use Safety Information Warnings and Cautions Read ALL Instructions Before UsingExplanation of Abbreviations Blender Specifications ActivationPressure Drop Transport / Storage RequirementsDryness and Composition for inlet gases Diagrams Component Description DescriptionPre-use testing Alarm TestReverse Gas Flow Procedure Operating InstructionsCleaning MaintenanceSettings Flowmeter Oxygen flowmeter conversionsTechnical Description Returns Disposal Instructions Troubleshooting Problem Probable Cause RemedyLimited Warranty Limitation of Liability Two 2 years from shipmentDeclaration of Conformity