Specifications continued
Dimensions: (without fittings)
Depth: | 4.9 in | (12.5 cm) |
Width: | 2.3 in | (5.7 cm) |
Height: | 4.1 in | (10.4 cm) |
Weight: | 2.9 lbs | (1.3 kg) |
Shipping Weight: | 3.5 lbs | (1.6 kg) |
Operating Temperature Range: | 59°F to 104°F (15°C to 40°C) |
Transport / Storage Requirements
Temperature Range: | |
Humidity: | Max 95% Noncondensing |
F O Accuracy:* | ± 3% of full scale @ 50 psi (3.4 bar) |
I 2 |
Pressure Drop:
Low Flow: ≤ 2 psi (0.14 bar) at inlet pressures from
(2.1- 6.2 bar) and at 10 lpm flow rate at 60% FIO2.
High Flow: ≤ 3 psi (0.21 bar) at inlet pressures from
(2.1- 6.2 bar) and at 30 lpm flow rate at 60% FIO2.
The
The
The alarmed Oxygen Monitor / Analyzer should comply with ISO 7767 to meet CE requirement.
Dryness and Composition for inlet gases:
Heliox: Medical grade 80/20 or 70/30 is required.
Oxygen: Oxygen supply must meet all requirements of USP Medical Oxygen.
Dew Point:
(ONLY for CE requirements)
Both inlets should remain 10°F
* Accuracy of FIO2 will be affected if bleed flow is not engaged at low flows. (At or below 3 lpm for Low Flow and 15 lpm for High Flow).
*When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow to maintain accuracy.
Specifications are subject to change without prior notice.
HELIO2 |
|
Blender | 5 |