IBM 1000-90 LI-1400 DataLogger, LI-1200S Minimum Data Set Recorder

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
Download on canonical page 22 pages, 124.78 Kb
8 COMM.EXE
The data dump can be aborted by pressing the OUT key again.
LI-1400 DataLogger
The LI-1400 is configured as a DTE device. Most computers can be
connected to the LI-1400 using the 1400-550 printer cable. To send
files from the LI-1400 to a computer, follow these steps:
1. Use the Setup Hardware menu to set the Printing baud rate to
9600.
2. Use the Setup Logging menu to make sure that Auto Print=No.
3. Press the Fct key.
4. Press the right or left arrow key until 'Print Memory' is
displayed. Press Enter.
5. Press the right or left arrow key until 'Print=All' is displayed.
6. Press the down arrow key several times until !Print Mem! is
displayed. Press Enter. The data will be sent.
LI-1200S Minimum Data Set Recorder
The LI-1200S is configured as a DTE device. Most computers can
be connected to the LI-1200S using the 1000-03 printer cable and a
female-to-female gender changer. To send files from the LI-1200S
to a computer, follow these steps:
1. Press OUT to enter Output Mode.
2. Choose baud rate (300 or 4800).
3. Press ENTER to begin data transmission.
MENU