Page 3
PERFORMANCE VERIFICAITONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed.
Performance Verification Preparation
1.Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization should be reached after 30 minutes.
2.Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.
3.Input the values listed in Table 1 for the Standard
Table 2 Performance Verification
|
|
|
|
Function Setting /Range | Input | Low Limit | High Limit |
ACV 3.2V | 3.000V @ 50Hz | 2.936 | 3.064 |
ACV 3.2V | 3.000V @ 500Hz | 2.936 | 3.064 |
ACV 32 | 30.00 @ 50Hz | 29.36 | 30.64 |
ACV 32 | 30.00 @ 500Hz | 29.36 | 30.64 |
ACV 320 | 300V AC @ 50Hz | 293.6 | 306.4 |
ACV 320 | 300V AC @ 500Hz | 293.6 | 306.4 |
ACV 600 | 500V AC @ 50Hz | 486 | 514 |
ACV 600 | 500V AC@ 60Hz | 486 | 514 |
DCV 320mV | 300mV DC | 296.3 | 303.7 |
DCV 3.2V | 3.000V DC | 2.975 | 3.025 |
DCV 32V | 30.00V DC | 29.63 | 30.37 |
DCV 320 | 300.0V DC | 296.3 | 303.7 |
DCV 600 | 500V DC | 493 | 507 |
A DC 320µA | 300µA | 293.9 | 306.1 |
A DC 3200 µA | 3000µA | 2939 | 3061 |
A DC 32mA | 30mA | 29.39 | 30.61 |
A DC 320mA | 300mA | 293.9 | 306.1 |
A DC 10 | 9.00 DCA | 8.70 | 9.30 |
A AC 320µA | 300µA @ 50Hz | 292.1 | 307.9 |
A AC 320µA | 300µA @ 500Hz | 292.1 | 307.9 |
A AC 3200µA | 3000µA @ 50Hz | 2921 | 3079 |
A AC 320oµA | 3000µA @ 500Hz | 2921 | 3079 |
A AC 32mA | 30mA @ 50Hz | 29.21 | 30.79 |
A AC 32mA | 30mA @ 500Hz | 29.21 | 30.79 |
A AC 320mA | 300mA @ 50Hz | 292.1 | 307.9 |
A AC 320mA | 300mA @500Hz | 292.1 | 307.9 |
A AC 10A | 9A @ 50Hz | 8.64 | 9.36 |
A AC 10A | 9A @ 500Hz | 8.64 | 9.36 |
Ω 320 | 300.0 | 297.7 | 306.3 |
Ω 2k | 3.000k | 2.950 | 3.048 |
Ω 32k | 30.00k | 29.52 | 30.48 |
Ω 320k | 300.0k | 295.2 | 304.8 |
Ω 2M | 3.000M | 2.922 | 3.078 |
Ω 32M | 30.00M | 28.45 | 31.55 |
Diode Test | 500mV | .448 | .552 |
Continuity Test | 120Ω | Beep off |
|
Continuity Test | 80Ω | Beep on |
|
|
|
|
|
Form number 61362 | rev 1 May 2002 |