Page 3

PERFORMANCE VERIFICAITONS

Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning correctly. If the meter does not conform to any of the listed limits the calibration procedure must be performed.

Performance Verification Preparation

1.Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization should be reached after 30 minutes.

2.Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.

3.Input the values listed in Table 1 for the Standard 61-362

Table 2 Performance Verification 61-362

 

 

 

 

Function Setting /Range

Input

Low Limit

High Limit

ACV 3.2V

3.000V @ 50Hz

2.936

3.064

ACV 3.2V

3.000V @ 500Hz

2.936

3.064

ACV 32

30.00 @ 50Hz

29.36

30.64

ACV 32

30.00 @ 500Hz

29.36

30.64

ACV 320

300V AC @ 50Hz

293.6

306.4

ACV 320

300V AC @ 500Hz

293.6

306.4

ACV 600

500V AC @ 50Hz

486

514

ACV 600

500V AC@ 60Hz

486

514

DCV 320mV

300mV DC

296.3

303.7

DCV 3.2V

3.000V DC

2.975

3.025

DCV 32V

30.00V DC

29.63

30.37

DCV 320

300.0V DC

296.3

303.7

DCV 600

500V DC

493

507

A DC 320µA

300µA

293.9

306.1

A DC 3200 µA

3000µA

2939

3061

A DC 32mA

30mA

29.39

30.61

A DC 320mA

300mA

293.9

306.1

A DC 10

9.00 DCA

8.70

9.30

A AC 320µA

300µA @ 50Hz

292.1

307.9

A AC 320µA

300µA @ 500Hz

292.1

307.9

A AC 3200µA

3000µA @ 50Hz

2921

3079

A AC 320oµA

3000µA @ 500Hz

2921

3079

A AC 32mA

30mA @ 50Hz

29.21

30.79

A AC 32mA

30mA @ 500Hz

29.21

30.79

A AC 320mA

300mA @ 50Hz

292.1

307.9

A AC 320mA

300mA @500Hz

292.1

307.9

A AC 10A

9A @ 50Hz

8.64

9.36

A AC 10A

9A @ 500Hz

8.64

9.36

Ω 320

300.0

297.7

306.3

Ω 2k

3.000k

2.950

3.048

Ω 32k

30.00k

29.52

30.48

Ω 320k

300.0k

295.2

304.8

Ω 2M

3.000M

2.922

3.078

Ω 32M

30.00M

28.45

31.55

Diode Test

500mV

.448

.552

Continuity Test

120Ω

Beep off

 

Continuity Test

80Ω

Beep on

 

 

 

 

 

Form number 61362

rev 1 May 2002

Page 5
Image 5
3Com 61-362 technical manual Performance Verificaitons, Performance Verification Preparation