1
Up and Running
Up & Running
Up and Running
1.Make sure the wet bulb’s wick is clean. Fill the reservoir with distilled water.
2.Place the QUESTemp in the work area in a safe location approximately 3.5 feet off the ground.
3.Turn the unit ON. If the battery voltage displayed during the
4.Use the arrow keys to set the display to the desired items.
5.Allow 10 minutes for the sensors to stabilize to the environment before taking readings.
Check Wick and Fill Natural Wet Bulb
The QUESTemp uses a cotton wick immersed into a reservoir containing distilled water. Ordinary tap water should not be used, as the contaminants that are left behind after evaporation will shorten the life of the wick and cause high readings. If the wick is discolored it should be replaced. To replace the wick, slide the old wick off the top of the sensor. Place a new wick over the sensor, making sure that the bottom of the wick is down in the reservoir.
Wick
Reservoir cover Reservoir
Filling wet bulb reservoir
1.Remove reservoir cover and fill with distilled or
2.Replace reservoir cover.