Advanced Manual Inflate UA-705 Blood Pressure Monitor

To obtain an accurate reading, it is important to use the correct cuff size. A cuff that is too small will result in a higher than expected reading and a cuff that is too large will result in a lower than expected reading. The Advanced Manual Inflate unit comes in two different cuff sizes to accommodate your customer’s needs. This monitor is not compatible with the LifeSource small cuff.

UNITED STATES

CANADA

 

PART

UPC

 

CUFF

 

 

NUMBER

NUMBER

 

SIZE

 

 

UA-705V

0-93764-60157-6

Medium 9.4 - 14.2” (24 - 36 cm)

 

 

 

 

UA-705VL

0-93764-60158-3

Large

14.2 - 17.7” (36 - 45 cm)

 

 

 

 

 

 

 

UA-705CN

0-93764-60159-0

Medium 9.4 - 14.2” (24 - 36 cm)

 

 

UA-705CNL

0-93764-60160-6

Large

14.2 - 17.7” (36 - 45 cm)

 

 

 

 

 

 

 

Each unit comes with a trilingual instruction manual, quick start card, AA battery, four week blood pressure logbook and warranty card.

Accessories

PART #

DESCRIPTION

UA-700

Gray Latex Free Inflator Bulb

PA:4000735

Exhaust Valve

UA-280

Medium Blood Pressure Monitor Cuff: 9.4” - 14.2” (24-36cm)

UA-281

Large Blood Pressure Monitor Cuff: 14.2” - 17.7” (36-45cm)

Technical Specifications

Dimensions

3.2” (w) x 4.2” (d) x 2.2” (h) (81mm x 105mm x 54mm)

Weight

4.3 oz. (120g) without batteries

Display

Digital: Pressure - 15mm character height; Pulse - 10mm character height.

 

Pressure and pulse displayed simultaneously.

Pressurization

Manual, by inflator bulb

Deflation

Hand-held push button

Type

Oscillometric

Accuracy

Pressure: ±3mmHg or 2%, whichever is greater/Pulse: ±5%

Pressure Range

20mmHg to 280mmHg/Pulse: 40 to 200 pulses/minute

Power Source

One type AA alkaline batteries (1.5 volt). Battery included.

Battery Life

Approx. 2,000 measurements

Operating Environ.

50°F to 104°F (10°C to 40°C)

Storage Environ.

14°F to 140°F (-10°C to 60°C)

Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute (ANSI/AAMI SP10) for electronic sphygmomanometers. Manufactured by an ISO 9000 and ISO 13485 registered company. Engineered in Japan. Made in China.

Packaging and Shipping Specifications

Selling Unit (Box)

W x D x H (inches)

WT (lb.)

Individual Item Cube (cubic feet)

All UA-705 models

5 1/16 x 3 1/2 x 7 11/16

1.32

0.08

Unit Case (10 units)

W x D x H (inches)

WT (lb.)

Case Cube (cubic feet)

All UA-705 models

18 1/2 x 11 x 8 1/2

14.34

1.00

Ask your A&D Sales Representative about point-of-purchase floor and counter displays available for this blood pressure monitor unit.

 

 

8/05

1756 Automation Parkway

LIT:LS705

San Jose, CA 95131

 

www.LifeSourceOnline.com

 

Pressure Rating Indicator is a trademark of A&D Medical. Patent Pending. LifeSource is a registered trademark of A&D Medical.

 

©2005 A&D Medical. All rights reserved.

 

Ask your A&D Sales Representative about other LifeSource products.

Blood Pressure Monitors

Digital Thermometers

Personal Scales

Stethoscopes

For questions about LifeSource products, call us toll-free at 1-800-726-3364

or visit our web site at www. LifeSourceOnline.com

Page 2
Image 2
A&D ua-705 quick start Accessories, Technical Specifications, Packaging and Shipping Specifications