EMC Tables

Recommended separation distances between portable and mobile RF communications equipment and the UT-302PBT Thermometer

The UT-302PBT Thermometer monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the UT-302PBT Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the UT-302PBT Thermometer as recommended below, according to the maximum output power of the communications equipment.

Rated maximum

Separation distance according to frequency of transmitter

output power of

 

m

 

transmitter

150 kHz to 80 MHZ

80 MHz to 800 MHZ

800 MHz to 2.5 GHz

W

d=1.2 √ P

d=1.2 √ P

d=1.2 √ P

 

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.