EMC Tables

Guidance and manufacturer's declaration-electromagnetic emission

The UT-302PBT Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the UT-302PBT Thermometer should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment guidance

 

 

 

RF emissions

CISPR 11

Group 1 The UT-302PBT Thermometer uses RF energy only

 

 

for its internal function. Therefore, its RF emissions are very

 

 

low and are not likely to cause any interference in nearby

 

 

electronic equipment.

RF emissions

Class B

 

CISPR 11

 

 

Harmonic

Not

The UT-302PBT Thermometer is suitable for use in all

emissions

applicable

establishments, including domestic establishments and

IEC 61000-3-2

 

those directly connected to the public low-voltage power

Voltage

Not

supply network that supplies buildings used for domestic

purposes.

fluctuations/

applicable

flicker emissions

 

 

IEC 61000-3-3

 

 

 

 

 

Guidance and manufacturer's declaration-electromagnetic immunity

The UT-302PBT Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the UT-302PBT Thermometer should assure that it is used in such an environment.

Immunity test

IEC 60601

Compliance

Electromagnetic environment-guidance

test level

level

 

 

Electrostatic

± 6 kV

± 6 kV

Floors should be wood, concrete or ceramic

discharge (ESD)

contact

contact

tile. If floors are covered with synthetic

IEC 61000-4-2

± 8 kV air

± 8 kV air

material, the relative humidity should be at

 

 

 

least 30 %.

 

 

 

 

Power frequency

3 A/m

3 A/m

Power frequency magnetic fields should be at

(50/60 Hz)

 

 

levels characteristic of a typical location in a

magnetic field IEC

 

 

typical.

61000-4-8

 

 

 