Agilent 34970A/Option 001
Internal DMM Assembly
Installation Instructions for 34970-80010 Retrofit Kit
1Disconnect the power cord.
2Remove the outer cover from the Agilent 34970A.
To remove the cover from the instrument, follow the procedures shown in chapter 6 of the 34970A Service Guide.
3Connect the internal DMM ribbon cable to the main circuit board.
Connect the loose end of the grey ribbon cable to connector J202 on the 34970A main circuit board
Snap the Locking Clips over the top of the connector to secure the Ribbon Cable
Ribbon Cable
1