Agilent 34970A/Option 001

Internal DMM Assembly

Installation Instructions for 34970-80010 Retrofit Kit

1Disconnect the power cord.

2Remove the outer cover from the Agilent 34970A.

To remove the cover from the instrument, follow the procedures shown in chapter 6 of the 34970A Service Guide.

3Connect the internal DMM ribbon cable to the main circuit board.

Connect the loose end of the grey ribbon cable to connector J202 on the 34970A main circuit board (34970-66511). Press down firmly on the ribbon cable connector and secure the cable in place with the locking clips.

Snap the Locking Clips over the top of the connector to secure the Ribbon Cable

Ribbon Cable

1

Page 1
Image 1
Agilent Technologies 34970-80010 installation instructions Agilent 34970A/Option Internal DMM Assembly