9Modify the Labeling on the Rear Panel.
If you originally ordered the 34970A without the internal DMM, a solid black circle should be visible above the
| Internal DMM |
2SW | is NOT installed. |
| |
|
|
| Internal DMM |
2SW | is installed. |
| |
|
|
10Replace the outer cover on the Agilent 34970A.
To replace the cover on the instrument, follow the procedures shown in chapter 6 of the 34970A Service Guide.
Calibration Information
The internal DMM was calibrated before it was shipped from the factory (a calibration certificate is included). However, after installing the internal DMM, you will want to make sure that you run the Zero Adjustment procedure as outlined in chapter 4 in the 34970A Service Guide.
The internal DMM is guaranteed to meet the instrument’s