1510

OUTPUT @Ps;"DIAG:EEPR ’’;Addr(I);’’,’’;Length(I);’’,’’;Init_data(I)

1520

NEXT I

 

 

 

1530

GOTO Cal_off

 

 

 

1540

!

 

 

 

1550

Fact_preset:

 

!

 

1560

CLEAR SCREEN

 

 

1570

PRINT "This program should ONLY be completed if your power supply”

1580

PRINT "EEPROM has been replaced or a component that will effect"

1590

PRINT "the calibration AND the alignment of voltage, overvoltage"

1600

PRINT "and current is complete AND unit has passed the performance"

1610

PRINT "test.

 

Enter C to continue, any other key to abort.”

1620

INPUT Cont_prog$

 

1630

IF (UPC$(Cont_prog$))< >"C" THEN GOTO Cal_off

 

1640

!

 

 

 

1650

CLEAR SCREEN

 

 

1660

PRINT "Transferring calibration data to factory preset locations."

1670

!

 

 

 

1680

Fact_cal_sour:

 

! Address of factory calibration data source

1690

DATA 2,6,62,66,20,24,70,74

 

1700

!

 

 

 

1710

Fact_cal_dest

:

! Address of factory calibration data destination

1720

DATA 78,82,86,90,94,98,102,106

 

1730

!

 

 

 

1740

Fact_cal_len:

 

! Length of factory calibration data

 

1750

DATA 4,4,4,4,4,4,4,4,

 

1760

!

 

 

 

1770

RESTORE Fact_cal_sour

 

1780

FOR I=1 TO 8

 

 

 

1790

READ Cal_sour_addr(I)

 

1800

NEXT I

 

 

 

1810

!

 

 

 

1820

RESTORE Fact_cal_dest

 

1830

FOR I=1 T0 8

 

 

 

1840

READ Cal_dest_addr(I)

 

1850

NEXT I

 

 

 

1860

!

 

 

 

1870

FOR I=1 T0 8

 

 

! Locations of good data

1880

OUTPUT @Ps;"DIAG:EEPR? ";Cal_sour_addr(I);",";4

! Read good data

1890

ENTER @Ps;Cal_data$

! Enter good data

1900

OUTPUT @Ps;"DIAG:EEPR";Cal_dest_addr(I);",";4;",";Cal_data$ ! Write data

1910

NEXT I

 

 

 

1920

!

 

 

 

1930

Cal_off !

 

 

 

1940

CLEAR SCREEN

 

 

1950

OUTPUT @Ps;"CaL:STATE OFF"

! Turn off cal mode

1960

!

 

 

 

1970

GOSUB Ps_error

 

! Check for errors

1980

IF Err THEN

 

 

 

1990

PRINT "An error occurred during the EEPROM read/write, Check for"

2000

PRINT "programming errors. Initialization data may be incorrect."

Figure 3-14. EEPROM Initialization Program (Sheet 4 of 5)

Troubleshooting 55

Page 55
Image 55
Agilent Technologies 6610XA appendix Eeprom Initialization Program Sheet 4 Troubleshooting

6610XA specifications

Agilent Technologies has long been recognized for its innovative contributions to the fields of measurement and testing, and the Agilent 6610XA series of power supplies is no exception. Tailored for demanding applications in electronics and advanced research, the 6610XA series exemplifies cutting-edge technology blended with user-friendly features.

One of the main characteristics of the Agilent 6610XA is its ability to provide a stable, accurate, and reliable power supply across a range of applications. With output voltages that can reach as high as 60V and currents up to 10A, the device is versatile enough to cater to a variety of testing requirements. This power supply is ideal for applications including semiconductor testing, device characterization, and more, where precision and consistency are paramount.

Among the standout features of the Agilent 6610XA is its advanced graphical user interface (GUI), which enhances the overall user experience. The intuitive design allows engineers and technicians to monitor and control voltage and current settings easily, making the process of configuring the device both fast and efficient. Additionally, the device has built-in measurement capabilities that can display real-time voltage and current readings, significantly aiding in troubleshooting and performance evaluation.

Furthermore, the 6610XA incorporates a range of communication interfaces including USB, LAN, and GPIB, making it highly adaptable for integration into various automated test setups. This versatility signifies that the power supply can be seamlessly incorporated into existing laboratory environments, promoting productivity and efficiency.

The series also incorporates intelligent protection mechanisms to ensure both user safety and equipment longevity. Features such as overvoltage protection (OVP) and overcurrent protection (OCP) are designed to prevent accidental overloads, safeguarding both the device under test and the power supply itself.

Another key aspect is the series' capability to perform complex programming tasks with ease. With programming capabilities that enable users to set intricate voltage and current profiles, the device supports advanced applications, including load transient testing and sweep testing. This flexibility makes the Agilent 6610XA a valuable asset for any research and development environment looking to enhance testing efficiency and accuracy.

In conclusion, the Agilent 6610XA series power supply stands out for its combination of precision, user-friendliness, advanced communication capabilities, and safety features. These attributes make it a critical tool for engineers and researchers engaged in the rigorous demands of modern electronics testing and evaluation. With its continued commitment to innovation, Agilent Technologies reinforces its position as a leader in providing high-quality solutions for the measurement and testing industry.