Verification and Calibration for Option 019

Verification and Calibration for Option 019

There are no changes to the output verification and calibration sections of the User's Guide for units with Option 019.

Option 019 Factory Modifications

Make the following changes to the Agilent 6813B Service Manual (p/n 5962-0859) for Agilent 6813B Option 019 units.

Initialization

For ROM revision A.00.07, change step 3 to “Scroll to 6813J02 and press Enter”.

NOTE: Option J02 and Option 019 are identical; the J02 option was renamed option 019. After an Agilent 6813B Option 019 Rev A.00.07 has been re-initialized, the number <1002> will be returned by a *OPT? query.

For ROM revision A.00.08 or higher, change step 3 to “Scroll to 6813B19 and press Enter”.

NOTE: An Agilent 6813B Option 019 Rev A.00.08 will return <019> for a *OPT? Query.

Replaceable Parts

Add or change the following part numbers in the parts list:

Reference Desig.

Part Number

Description

A4

5064-0081

AC Input Assembly Tested

A4 C1060 - C1067

0180-4963

Capacitor 3300uF 200V

A9

5064-0083

Output/Relay/Filter Assembly Tested

A9 R552 – R553

0698-3156

Resistor 14.7K 1%

A10

5064-0079

Control Assembly Tested

 

5964-8149

Appendix Option 019

 

9320-5540

Label-Information