GENERAL INFORMATION

INTRODUCTION

This manual describes all models in the Agilent E361xA 60W Bench Power Supply family and unless stated otherwise, the information in this manual applies to all models.

SAFETY REQUIREMENTS

This product is a Safety Class I instrument, which means that it is provided with a protective earth ground terminal. This terminal must be connected to an ac source that has a 3-wire ground receptacle. Review the instrument rear panel and this manual for safety markings and instructions before operating the instrument. Refer to the Safety Summary page at the beginning of this manual for a summary of general safety information. Specific safety information is located at the appropriate places in this manual.

This power supply is designed to comply with the following safety and EMC(Electromagnetic Compatibility) require- ments:

nIEC 348: Safety Requirements for Electronic Measuring Apparatus

nIEC 1010-1/EN 61010: Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use

nCSA C22.2 No.231: Safety Requirements for Electrical and Electronic Measuring and Test Equipment

nUL 1244: Electrical and Electronic Measuring and Testing Equipment.

nEMC Directive 89/336/EEC: Council Directive entitled Approximation of the Laws of the Member States relating to Electromagnetic Compatibility

nEN 55011(1991) Group 1, Class B/CISPR 11: Limits and nMethods of Radio Interference Characteristics of nIndustrial, Scientific, and Medical(ISM) Radio-Frequency

Equipment

nEN 50082-1(1991) /

IEC 801-2(1991):Electrostatic Discharge Requirements IEC 801-3(1984):Radiated Electromagnetic Field

Requirements

IEC 801-4(1988):Electrical Fast Transient/Burst

Requirements

INSTRUMENT AND MANUAL IDENTIFICATION

A serial number identifies your power supply. The serial number encodes the country of manufacture, the date of the latest significant design change, and a unique sequential number. As an illustration, a serial number beginning with KR306 denotes a power supply built in 1993 (3=1 993, 4=1994, etc), 6th week manufacture in Korea(KR). The remaining digits of the serial number are a unique, five-digit number assigned sequentially.

If the serial number on your supply differs from that shown on the title page of this manual, a yellow MANUAL CHANGES sheet is supplied with this manual to explain the difference between your instrument and the instrument described by this manual. The change sheet may also con- tain information for correcting errors in the manual.

OPTIONS

Options OE3 and OE9 determine which line voltage is selected at the factory. The standard unit is configured for 115 Vac ± 10%. For information about changing the line voltage setting, see paragraph "INPUT POWER REQUIREMENTS", page 1-6.

OE3:

Input power, 230 Vac ± 10%, 47-63 Hz

OE9:

Input power, 100 Vac ± 10%, 47-63 Hz

910:One additional manual

ACCESSORY

The accessory listed below may be ordered from your local Agilent Technologies Sales Office either with the power sup- ply or separately. (Refer to the list at the rear of the manual for address.)

Agilent Part No. Description

5063-9240 Rack Kit for mounting one or two 3 1/2" high supply in a standard 19" rack

The rack mount kit is needed for rack mounting of all models in the Agilent E361xA power supply because these supplies have molded feet.

DESCRIPTION

This power supply is suitable for either bench or rack mounted operation. It is a compact, well-regulated, Constant Voltage/Constant Current supply that will furnish full rated output voltage at the maximum rated output current or can be continuously adjusted throughout the output range. The out- put can be adjusted both locally from the front panel and remotely by changing the settings of the rear panel switches (See paragraph "REMOTE OPERATING MODES", page 1-9). The models in this family offer up to 60 watts of output power, with voltage up to 60 volts and current up to 6 amps as shown in Table 1.

The front panel VOLTAGE control can be used to establish the voltage limit when the supply is used as a constant cur- rent source and the CURRENT control can be used to estab- lish the output current limit when the supply is used as a constant voltage source. The supply will automatically cross over from constant voltage to constant current operation and vice versa if the output current or voltage exceeds these pre- set limits.

The front panel includes an autoranging (E3614A single- range) digital voltmeter and a single-range digital ammeter. Two 3 1/2 digit voltage and current displays accurately show the output voltage and current respectively. The output rat- ings for each model are shown in the Specifications and Operating Characteristics Table.

The OVP/CC SET switch is used to check the OVP trip volt- age and current control set value. When pressing this switch, the voltage display indicates the OVP trip voltage and the cur- rent display indicates the current control set value.

The power supply has both front and rear output terminals. Either the positive or negative output terminal may be

1-4

Page 4
Image 4
Agilent Technologies Agilent E3616A Serials KR83502651, Agilent E3615A Serials KR83506197 service manual General Information

Agilent E3614A Serials KR83503035, Agilent E3617A Serials KR83502522, Agilent E3615A Serials KR83506197, Agilent E3616A Serials KR83502651 specifications

Agilent Technologies is a leading player in the field of electronic test and measurement devices, providing essential tools for engineers and scientists. Among their array of power supplies, models such as the Agilent E3616A, E3615A, E3617A, and E3614A stand out for their exceptional performance, reliability, and technological features.

The Agilent E3616A is a precision DC power supply known for its ability to deliver a maximum output of 60V and 3A. Its serial number, KR83502651, signifies its unique identity. This model is particularly valued in laboratory environments due to its advanced programming capabilities. It offers a comprehensive range of output voltage and current settings, making it versatile for various applications. One of its main features is the built-in 6V at 3A output, which can be useful for testing other devices.

Similarly, the Agilent E3615A, with the serial number KR83506197, provides an excellent blend of performance and functionality. It is capable of delivering up to 30V and 3A across its outputs. This model excels in its low noise operation, which is crucial for sensitive electronics testing. Furthermore, the user-friendly interface allows for quick adjustments and settings retrieval, making it a favorite among engineers. The E3615A is designed to support a wide range of applications, from component testing to powering devices in research settings.

The Agilent E3617A, identifiable by its serial number KR83502522, is another high-performance model that offers a maximum output of 30V and up to 3A. It features built-in read-back capabilities, enabling engineers to monitor the actual voltage and current supplied to the load. The E3617A stands out for its programmable output, allowing automated testing sequences to be implemented with ease.

Lastly, the Agilent E3614A, with the serial number KR83503035, is a cost-effective option without sacrificing quality. This model offers an output of up to 30V and 3A while maintaining stability and precision. Its compact design makes it easy to integrate into crowded workspaces or test setups. The E3614A is designed for durability, ensuring reliable performance over time, making it an excellent choice for both education and industrial applications.

With their high-precision outputs, easy programmability, and user-friendly interfaces, these Agilent power supplies serve as essential tools for a variety of testing and research needs, reflecting Agilent Technologies' commitment to quality and innovation in electronic measurement solutions.