Aiphone NH-1PA installation manual Nurse Center System, Wiring Diagram, Instructions

Models: NH-1PA

1 2
Download 2 pages 13.34 Kb
Page 1
Image 1
Nurse Center System
NH-1PA
(Underneath)
1 2 3 4
ON
OFF Value = 1 2 4 8

NHX

Nurse Center System

NH-1PA

Paging Adaptor for NHX System

Provides zone paging to an overhead speaker system

- INSTRUCTIONS -

Some applications for Nurse Call systems require paging to overhead speakers. The NH-1PA allows a button on the NHX master to be used for connection to an external paging amplifier. If multiple zones of paging are required, use one NH-1PA per zone.

NOTE: An NHX system installed with an NH-1PA included does not carry the UL 1069 listing.

WIRING DIAGRAM:

NHX-80XNH-1PA

PS-2420UL

+

A+

A

-

A-

G

 

 

+B+

- B-

PS-2420ul

Trunk Line

11

A1

Connection

12

A2

(Connect to trunk

13

P3

 

corresponding to

14

P4

Page Button

15

D5

location desired.

16

D6

 

 

 

PG-C Amplifier

3

OUTPUT

TO 70V

4

PAGING

 

SPEAKERS

 

 

AUX input

NOTE:

 

Aiphone amplifiers are available with outputs

 

 

of 20, 50 and 100 watts. If connecting to

 

another brand of amplifier, use AUX input.

 

NH-1PA (Back)

 

 

A1 A2

P3 P4

 

D5 D6

A G

NOTES:

1.When multiple NH-1PA's are included in the same system, each NH-1PA occupies a station button on the master station. This reduces the overall capacity for patient stations on the system.

2.The total amount of wire in a trunkage line including NH-1PA must not exceed 330'.

3.Adjust the volume of the paging speakers first from the volume control on the NH-1PA. Then adjust the volume control on the amplifier.

4.A system including zone paging does not carry the UL 1069 listing.

5.Only wiring concerning overhead paging is shown here. For complete installation and wiring information, please see the complete NHX Installation Manual.

CONTROLS AND SETTINGS:

DIP SWITCH SETTINGS (Located on bottomside of NH-1PA): Switch Values from left to right: 1-2-4-8

1. Set dip switches to the numerical value of the station button (1~10) to be designated as the PAGING button.

2. Station numbers are 1-10 vertically on the master station.

If the Paging button is the TOP button on a column (1), set dip switch 1 to ON. If the Paging button is the BOTTOM button (10), set switches 2 & 8 to ON.

3.Aiphone recommends setting the paging adaptor as station #10 on the first column (trunk line 1, wired as shown here.)

However, any station button can be designated for paging. The paging button is determined by (a) which trunk line the NH-1PA is wired to, and (b) the station number setting on the NH-1PA.

PAGING VOLUME (Located on front of NH-1PA):

This controls the overall output volume of the paging. Adjust to the desired level.

Page 1
Image 1
Aiphone NH-1PA installation manual Nurse Center System, Wiring Diagram, Paging Adaptor for NHX System, Instructions