Bacharach 50 Specifications, Operation, Turning the Snifit On/Off, Zeroing the Sensor, S n if it

Models: 50

1 6
Download 6 pages 47.64 Kb
Page 2
Image 2
Specifications

Specifications

Gas Monitored

Carbon Monoxide (CO)

Range

0 to 1999 ppm

Resolution

1 ppm

Accuracy

± (5% of reading + 5 ppm)

Response

≤30 sec. to 90% of final value

Power

9V battery

Battery Life

1500 hours minimum

Temperature Range ....

32 to 104 °F (0 to 40 °C)

Case Material

High impact plastic

Weight

0.25 lb (0.11 kg) without battery

Dimensions

5.5" L x 2.0" W x 1.6" H

 

(140 x 51 x 41 mm)

Operation

Turning the Snifit On/Off

If not already done, install a 9V battery as described under Battery Installation.

Turn on the instru- ment by pressing its Power On button.

Turn it off by pressing Power Off (or the instrument will automatically shut itself off after approxi- mately 35 minutes).

Z E

R

O

 

 

A D J U S T

 

 

 

 

 

 

S n if it

 

 

 

 

C A R B O N

 

 

 

 

M O N O X I D E A N A L Y Z

 

 

 

 

 

 

 

 

 

S P A N

 

 

 

 

m o d e lS p a n5

 

 

 

A D J U S T

L O W

 

A0

 

 

 

 

 

 

P O T .

B A T T E

R Y

 

 

 

 

 

 

L O B A T

 

 

I N D I C A

T I O N

 

 

D I S P L A

 

 

 

 

 

B a c k l i gOhft-fP o wOenr -

 

 

B A C K

L

I G H T

 

 

 

 

P O W E R

 

 

O N / O F F

 

 

 

 

 

 

 

 

 

 

 

P O W E R

 

 

 

 

p p m

 

C O

Figure 1. Snifit 50

Zeroing the Sensor

Turn on the Snifit and allow the displayed reading to stabilize before proceeding – approximately 30 sec- onds. Then with the instrument sampling fresh air

- 2 -

Page 2
Image 2
Bacharach 50 manual Specifications, Operation, Turning the Snifit On/Off, Zeroing the Sensor, S n if it, p p m