Calibration

Important! For accurate Snifit operation, periodic calibration of its CO sensor is required.

To calibrate the CO sensor, you will need the equip- ment listed under Accessories. Note that the Zero Gas Cylinder is needed only if you’re unsure about the quality of the surrounding air for zeroing purposes.

Calibrate the sensor to a known concentration of CO gas as follows:

1.Assemble the calibration equipment per Figure 2.

2.Zero the instrument as previously described under Zeroing the Sensor.

If necessary, you can use the calibration cup to apply a blend of Oxygen/Nitrogen gas directly over the sensor by attaching a zero gas cylinder to the regulator and adjusting the regulator knob for a flow rate of 2 SCFH.

3.Push the Calibration Cup over the sensor housing.

4.Attach a CO Span Gas Cylinder to the regulator. Then apply span gas to the sensor by adjusting the regulator knob for a flow rate of 2 SCFH.

-4 -