Calibration
Important! For accurate Snifit operation, periodic calibration of its CO sensor is required.
To calibrate the CO sensor, you will need the equip- ment listed under Accessories. Note that the Zero Gas Cylinder is needed only if you’re unsure about the quality of the surrounding air for zeroing purposes.
Calibrate the sensor to a known concentration of CO gas as follows:
1.Assemble the calibration equipment per Figure 2.
2.Zero the instrument as previously described under Zeroing the Sensor.
If necessary, you can use the calibration cup to apply a blend of Oxygen/Nitrogen gas directly over the sensor by attaching a zero gas cylinder to the regulator and adjusting the regulator knob for a flow rate of 2 SCFH.
3.Push the Calibration Cup over the sensor housing.
4.Attach a CO Span Gas Cylinder to the regulator. Then apply span gas to the sensor by adjusting the regulator knob for a flow rate of 2 SCFH.