Braun 6013, 6012 manual Product specifications, Accuracy for patient temperature range

Models: 6013

1 16
Download 16 pages 34.32 Kb
Page 15
Image 15
Product specifications

Product specifications

Displayed temperature range

93.2 °F -108 °F (34 °C - 42.2 °C)

Operating ambient

 

 

temperature range

50 °F -104 °F

(10 °C - 40 °C)

Display resolution

0.1 °F or °C

 

Accuracy for patient temperature range

 

 

Maximum Laboratory Error

96.8 °F -102.2 °F (36 °C - 39 °C):

± 0.4 °F

± 0.2 °C

outside this range:

± 0.5 °F

± 0.3 °C

Long term storage ranges

 

 

Temperature:

- 4 °F to 122 °F ( - 20 °C to 50 °C)

Humidity:

95 % non-condensing

Battery life:

2 years / 1000 measurements

This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the thermometer system [thermometer with lens filter]). Full responsibility for the conformance of the product to the standard is assumed by Braun GmbH, 61476 Kronberg, Germany.

ASTM laboratory accuracy requirements for the thermometer only (not including a lens filter) in the display range of 98.8 °F to 102.2 °F

(36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).

This appliance conforms to the following standards: DIN EN 60601-1: 3/96 «Medical electrical equipment» – Part 1: General requirements for safety

DIN EN 12470-5: 2003 «Clinical thermometers» –

Part 5: Performance of infrared ear thermometers (with maximum device)

TYPE BF

Attention, consult

MEDICAL ELECTRICAL EQUIPMENT

EQUIPMENT

ACCOMPANYING

CLASSIFIED BY UNDERWRITERS LABORATORIES INC.®

 

DOCUMENTS

WITH RESPECT TO ELECTRIC SHOCK, FIRE AND

 

 

MECHANICAL HAZARDS ONLY

 

 

IN ACCORDANCE WITH UL 2601-1 / CAN CSA C22.2

 

 

No.601.1 10HK

Internally Powered Equipment

Continuous Operation

Not Protected against Ingress of Water

U.S. Patent No. 5,088,834 Other Patents Pending

MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC requirements please contact the Braun ThermoScan Customer Service at 1-800-327-7226.

Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.

In the unlikely event you experience any difficulty using the Braun ThermoScan thermometer, simply call us toll free at 1-800-327-7226.