Troubleshooting
Error message | Situation | Solution |
| No lens filter is attached. | Attach new, clean lens filter |
No lens filter is attached and activation button was pressed while probe was in the ear.
Make sure probe tip is clean;
refer to section «Care and cleaning» if necessary.
Attach new, clean lens filter to stop error beeps.
|
| Ambient temperature is not | Allow the thermometer to remain | |
|
| within the allowed operating | in a room for 30 minutes where | |
|
| range 50 °F to 104 °F | the temperature is between | |
|
| (10 °C - 40 °C) or changing | 50 °F to 104 °F (10 °C - 40 °C) | |
|
| too rapidly. |
| |
|
|
|
| |
|
| Temperature taken is not | Make sure new, clean lens filter is | |
| ||||
|
| within typical human | attached and thermometer is properly | |
|
| temperature range (93.2 °F - | inserted. | |
|
| 108 °F / 34 °C - 42.2 °C). | Then, take a new temperature. | |
|
| HI = | too high |
|
|
|
| ||
|
| LO = | too low |
|
|
|
|
| |
|
|
|
|
|
|
| Battery is low, but | Insert new batteries | |
|
| |||
|
| thermometer will still operate |
| |
|
| correctly. |
| |
|
| (No light function on |
| |
|
| IRT 3020 models.) |
| |
|
|
| ||
|
|
|
| |
|
| Battery is too low to take | Insert new batteries | |
|
| |||
|
| correct temperature |
| |
|
| readings. |
| |
|
|
|
| |
|
|
|
|
|
|
| System error | Wait 2 minutes until thermometer | |
|
| |||
|
|
|
| powers down, then turn on again. |
|
| • If error persists | ... reset the thermometer by removing | |
|
|
|
| the batteries and putting them back in. |
|
| • If error still persists | ... call Braun ThermoScan Customer | |
|
| |||
|
|
|
| Service at 1- 800 - |