Braun 6012, 6013 manual Troubleshooting, Error message

Models: 6013

1 16
Download 16 pages 34.32 Kb
Page 12
Image 12
Troubleshooting

Troubleshooting

Error message

Situation

Solution

 

No lens filter is attached.

Attach new, clean lens filter

No lens filter is attached and activation button was pressed while probe was in the ear.

Make sure probe tip is clean;

refer to section «Care and cleaning» if necessary.

Attach new, clean lens filter to stop error beeps.

 

 

Ambient temperature is not

Allow the thermometer to remain

 

 

within the allowed operating

in a room for 30 minutes where

 

 

range 50 °F to 104 °F

the temperature is between

 

 

(10 °C - 40 °C) or changing

50 °F to 104 °F (10 °C - 40 °C)

 

 

too rapidly.

 

 

 

 

 

 

 

Temperature taken is not

Make sure new, clean lens filter is

 

 

 

within typical human

attached and thermometer is properly

 

 

temperature range (93.2 °F -

inserted.

 

 

108 °F / 34 °C - 42.2 °C).

Then, take a new temperature.

 

 

HI =

too high

 

 

 

 

 

 

LO =

too low

 

 

 

 

 

 

 

 

 

 

 

 

Battery is low, but

Insert new batteries

 

 

 

 

thermometer will still operate

 

 

 

correctly.

 

 

 

(No light function on

 

 

 

IRT 3020 models.)

 

 

 

 

 

 

 

 

 

 

Battery is too low to take

Insert new batteries

 

 

 

 

correct temperature

 

 

 

readings.

 

 

 

 

 

 

 

 

 

 

 

 

System error

Wait 2 minutes until thermometer

 

 

 

 

 

 

powers down, then turn on again.

 

 

• If error persists

... reset the thermometer by removing

 

 

 

 

the batteries and putting them back in.

 

 

• If error still persists

... call Braun ThermoScan Customer

 

 

 

 

 

 

Service at 1- 800 - 327-7226.

12

Page 12
Image 12
Braun 6012, 6013 manual Troubleshooting, Error message