CS510 OxyGuard Dissolved Oxygen Probe

 

 

 

The following instructions in program table 2 would trigger the agitator as

 

 

 

discussed in the agitator manual:

 

 

 

 

 

*Table 2 Program

 

 

 

02:

2.0

Execution Interval (seconds)

 

1: Do (P86)

 

 

 

 

1:

41

Set Port 1 High

; Use port number as wired

 

2: Excitation with Delay (P22)

 

 

1:

1

Ex Channel

 

 

2:

0

Delay W/Ex (units = 0.01 sec)

 

3:

25

Delay After Ex (units = 0.01 sec)

 

4:

0

mV Excitation

 

 

3: Do (P86)

 

 

 

 

1:

51

Set Port 1 Low

 

The above example is not as power efficient as possible, and would require AC power. If it is necessary to operate an agitator without AC power available, write the program so that the agitator is only operated for a short period of time just before the measurement is to be taken.

7