| CS510 OxyGuard Dissolved Oxygen Probe |
|
|
| The following instructions in program table 2 would trigger the agitator as | |
|
|
| discussed in the agitator manual: | |
|
|
|
| |
| *Table 2 Program |
|
| |
| 02: | 2.0 | Execution Interval (seconds) | |
| 1: Do (P86) |
|
|
|
| 1: | 41 | Set Port 1 High | ; Use port number as wired |
| 2: Excitation with Delay (P22) |
| ||
| 1: | 1 | Ex Channel |
|
| 2: | 0 | Delay W/Ex (units = 0.01 sec) | |
| 3: | 25 | Delay After Ex (units = 0.01 sec) | |
| 4: | 0 | mV Excitation |
|
| 3: Do (P86) |
|
|
|
| 1: | 51 | Set Port 1 Low |
|
The above example is not as power efficient as possible, and would require AC power. If it is necessary to operate an agitator without AC power available, write the program so that the agitator is only operated for a short period of time just before the measurement is to be taken.
7