CS510 OxyGuard Dissolved Oxygen Probe
3)Determine the air temperature and barometric pressure.
4)Using a calibration chart such as that provided in the probe’s manual, determine the oxygen concentration of the air.
5)Use the following equation to calculate the multiplier:
M = P/R
M = Multiplier
P = Concentration in PPM of the air (from the calibration chart)
R = The signal output of the OxyGuard probe when using a multiplier of one
6)Change the multiplier in the CR10 program from one to the calculated number.
10.Maintenance
The OxyGuard probe needs little maintenance. Regular cleaning of the membrane is all that is required. The membrane is very durable and can be cleaned with a cloth or soft paper. Do not scratch it clean with your fingernail. Please see the OxyGuard manual for complete maintenance procedures.
11. Agitator Control
In low flow conditions (less than about 2 cm/sec), it may be necessary to add an agitator to the OxyGuard probe. The distributor for OxyGuard in North America sells a
The wiring for the agitator as controlled by this example program would be as follows:
Ground/G |
|
|
|
| BLACK |
|
|
| 4 | 1 |
|
| Agitator | ||
Datalogger |
|
| #7321 Relay |
|
|
| |
|
|
|
|
|
| ||
|
|
| Crydom D1D07 |
|
|
|
|
Port C1 |
| 3 | 2 |
| WHITE |
| |
|
| ||||||
|
|
|
|
|
|
|
|
+-
12 V Battery
6