CS510 OxyGuard Dissolved Oxygen Probe
2. Specifications
Principle of Measurement: Membrane covered polarographic oxygen probe— self polarizing galvanic cell
Output Signal: 5 mV per mg/l, 2 mV per %/vol nominal
Accuracy: Better than +/- 2 % of reading +/- 1 digit when calibration temperature equals measuring temperature +/- 5 C
Repeatability: Better than +/- 0.2 mg/l; Better than +/- 0.5 %/vol when barometric pressure unchanged
Output Impedance: 1 kOhm nominal Response Time:
mg/l: 60 s for 90 % of 100 % step change at constant temperature
%vol: 10 s for 90 % of 100 % step change at constant temperature Materials of Construction:
Probe body: POM (polyoxymethylene)
Membrane: 0.05 mm (2 mil) Teflon Dimensions and Weight:
58mm diameter by 59 mm high
350g (3/4 lb) including standard 3 m (10 ft) cable length
Cable:
Operating Conditions: Temperature: 0 to 50 C (32 to 122 F)
Pressure: Maximum 10 atmospheres (147 psig)
Minimum Submersion Depth: 60 mm (2 ½ in)
Minimum Water Flow: 2.5 cm/s (1 in/sec) across membrane
Calibration: In air or in air saturated water. Use of EasyCal calibrator optional.
Temperature Compensation: Automatic from 4 - 40 C (40 - 104 F)
Probe Options: |
|
|
| Type I | Type II |
Range of dissolved | ||
oxygen |
|
|
Gaseous oxygen | ||
Electrode materials | Ag cathode/Zn anode | Ag cathode/Pb anode |
Probe electrolyte | NaCl | Na2CO3 |
2