Measurement system: Oscillometric method
Display Digital Display
Measuring Range Pressure 0 to 280 mmHg (in 1 mmHg increments)
Pulse 40 to 180 beats/min.
Accuracy Pressure ±3 mmHg
Pulse ±5% of reading
Infl ation Automatic Infl ation by Internal Pump
Infl ation setting Auto infl ation setting (140 to 250 mmHg)
Manual infl ation setting (160 to 280 mmHg)
Memory function Up to 7 memories
(Systolic blood pressure, Diastolic
blood pressure and pulse)
Decompression Constant exhaust valve system
Power Source Four “AA” batteries (R6P, LR6)
or optional AC/DC adopter (6V 4W)
Battery Life Alkaline batteries: Approx. 700 times
Manganese batteries: Approx. 200 times
(160 mmHg, 1 time/day, 22°C)
Rated Voltage DC 6V 4W ( :Direct Current)
Operating
Environment
Temperature 10 to 40°C
Humidity 30 to 85%
Storage Environment Temperature -20 to 60°C
Humidity 10 to 95%
Dimensions 130(H) x 95(W) x 118(D) mm
Weight 300g (Including Arm Cuff and Body without batteries)
Electrical shock protection Internal power unit
Degree of Protection Type BF applied part
Mode of operation: Continuous operation
Memory: 7 readings
Service Life: 5 years
Cuff’s Life: Approx. 2000 times
Accessories: Set includes an cuff (model: SCN-002), 4 AA batteries
(R6P) for the monitor, Instruction Manual
Optional Accessories: AC Adapter