Low Priority Alarms

Alarm Message

Description

Action

Replace C-filtr

Compressor filter needs

Replace filter. Press and

 

replacement.

hold mute key for more

 

 

than 4 seconds to reset

 

 

the hour counter.

 

 

 

Replace P-filtr

Patient filter needs

Replace filter. Press and

 

replacement.

hold mute key for more

 

 

than 4 seconds to reset

 

 

the hour counter.

Wash I-filtr

Input filter needs to be

 

cleaned.

Wash input filter. Press and hold mute key for more than 4 seconds to reset the hour counter.

 

Replace batt

Main battery has exceeded

Contact Customer

 

 

300 recharge cycles.

Service for replacement.

 

 

 

Note: This alarm will be

 

 

 

automatically cleared once

 

 

 

a new battery is inserted.

 

 

 

Battery will continue to

 

 

 

function after 300 recharge

 

 

 

cycles, but for a shorter time

 

 

 

than a new battery.

 

 

 

 

 

 

XX: Service*

Service required.

Contact Customer

 

 

 

Service.

 

 

 

 

 

 

*Value: 71-99

 

 

 

Other Messages

Message

Description

Action

 

 

 

Power removed

External power has been

No action is required.

 

disconnected; unit is now

 

 

running on battery.

 

 

 

 

SW Ver N.NNN

Display of software version

No action is required.

 

during startup and calibra-

 

 

tion. This includes test of

 

 

beeper and LEDs.

 

 

 

 

Self test

Indicates unit is going

No action is required.

 

through its test at startup.

 

 

Displayed for approximately

 

 

10 seconds.

 

 

 

 

Shutting off

Displayed while unit goes

No action is required.

 

through its power-down

 

 

sequence.

 

 

 

 

External power

External power is

No action is required.

 

connected.

 