Low Priority Alarms
Alarm Message | Description | Action |
Replace | Compressor filter needs | Replace filter. Press and |
| replacement. | hold mute key for more |
|
| than 4 seconds to reset |
|
| the hour counter. |
|
|
|
Replace | Patient filter needs | Replace filter. Press and |
| replacement. | hold mute key for more |
|
| than 4 seconds to reset |
|
| the hour counter. |
Wash | Input filter needs to be |
| cleaned. |
Wash input filter. Press and hold mute key for more than 4 seconds to reset the hour counter.
| Replace batt | Main battery has exceeded | Contact Customer | |
|
| 300 recharge cycles. | Service for replacement. | |
|
|
| Note: This alarm will be | |
|
|
| automatically cleared once | |
|
|
| a new battery is inserted. | |
|
|
| Battery will continue to | |
|
|
| function after 300 recharge | |
|
|
| cycles, but for a shorter time | |
|
|
| than a new battery. | |
|
|
|
|
|
| XX: Service* | Service required. | Contact Customer | |
|
|
| Service. | |
|
|
|
|
|
| *Value: |
|
|
|
Other Messages
Message | Description | Action |
|
|
|
Power removed | External power has been | No action is required. |
| disconnected; unit is now |
|
| running on battery. |
|
|
|
|
SW Ver N.NNN | Display of software version | No action is required. |
| during startup and calibra- |
|
| tion. This includes test of |
|
| beeper and LEDs. |
|
|
|
|
Self test | Indicates unit is going | No action is required. |
| through its test at startup. |
|
| Displayed for approximately |
|
| 10 seconds. |
|
|
|
|
Shutting off | Displayed while unit goes | No action is required. |
| through its |
|
| sequence. |
|
|
|
|
External power | External power is | No action is required. |
| connected. |
|