Other Messages, continued
Message | Description | Action |
Battery low | Estimated battery life less |
| than 17 minutes. The unit |
| will also generate a high |
| priority alarm as described |
| above. |
Charge the battery pack by plugging in to power supply.
Note: The message will be automatically cleared when plugged in to power supply.
Fully charged | Displayed as the battery | No action is required. |
| menu item when the |
|
| battery is 100% charged. |
|
|
|
|
Checking power | Displayed as the battery | No action is required. |
| menu item when the |
|
| battery status has recently |
|
| changed and the unit |
|
| software is analyzing the |
|
| change in status. |
|
No battery | Displayed as the battery |
| menu item when there are |
| no communications with |
| the battery. |
Verify that the battery pack is correctly installed. Contact Customer Service if the battery is fully inserted and the message continues to be displayed longer than 30 seconds.
|
| BattLife NN% | Displayed percentage of | Message is displayed | ||
|
|
| battery charge if at least | when mode key is | ||
|
|
| 10% and there is no | pressed. | ||
|
|
| external power connected. |
|
|
|
|
| Charging: NN% | NN% displays the cur- | Message is displayed | ||
|
|
| rent battery charge level. | when mode key is | ||
|
|
| Displayed when battery | pressed. | ||
|
|
| charge is greater than 10% |
|
|
|
|
|
| but less than 100% and |
|
|
|
|
|
| there is external power |
|
|
|
|
|
| connected. |
|
|
|
|
| Charging | Battery charge is less than | Message is displayed | ||
|
|
| 10% and there is external | when mode key is | ||
|
|
| power connected. | pressed. | ||
|
|
|
|
|
| |
|
| RT XXXXX | No action required. | |||
|
|
| time indicator (hours). |
|
|
|
|
|
|
|
|
| |
|
| Breath rate XX | The patient’s average breath | Reduce activity level. Be | ||
|
|
| rate when the device is | sure air inlet/outlet has | ||
|
|
| delivering the maximum | not been blocked. | ||
|
|
| amount of oxygen and | Note: The message will | ||
|
|
| the bolus is reduced. If no | |||
|
|
| automatically clear when | |||
|
|
| breaths are detected, the | the device returns to normal | ||
|
|
| most recent breath rate is | operation. | ||
|
|
| shown. |
|
|
|
|
|
|
|
|
| |
|
| Alarm cleared | A previously set alarm has | No action required. | ||
|
|
| been automatically cleared. |
|
|
|
|
|
|
| |||
|
|
|
|
|