Guidance and manufacturer’s declaration-electromagnetic immunity-

for all EQUIPMENT AND SYSTEMS

Guidance and manufacturer’s declaration-electromagnetic immunity

The ET1928L is intended for use in the electromagnetic environment specified below. The customer or the user of the ET1928L should assure that it is used in such an environment.

Immunity test

IEC 60601

Compliance level

Electromagnetic environment-

test level

 

 

guidelines

 

 

 

 

Electrostatic

± 6 kV contact

± 6 kV contact

Floors should be wood, concrete or

discharge(ESD)

± 8 kV air

± 8 kV air

ceramic tile. If floors are covered with

IEC 61000-4-2

 

 

synthetic material, the relative humidity

 

 

 

should be at least 30%.

 

 

 

 

Electrical Fast

± 2 kV for power

± 2 kV for power

Mains power quality should be that of a

transient/burst

supply lines

supply lines

typical commerical or hospital

 

 

 

environment.

IEC 61000-4-4

± 1 kV for input/output

± 1 kV for input/output

 

 

lines

lines

 

 

 

 

 

Surge

± 1 kV line(s) to line(s)

± 1 kV line(s) to line(s)

Mains power quality should be that of a

IEC 61000-4-5

± 2 kV line(s) to earth

± 2 kV line(s) to earth

typical commerical or hospital

 

 

 

environment.

 

 

 

 

Voltage dips, short

<5% UT

<5% UT

Mains power quality be that of a

interruption and

(>95% dip in UT)

(>95% dip in UT)

typical commerical or hospital

voltage variations

for 0.5 cycle

for 0.5 cycle

environment. If the user of the ET1928L

on power supply

 

 

requires continued operation during power

input lines

40% UT

40% UT

mains interruptions, it is recommended that

 

(60% dip in UT)

(60 % dip in UT)

the ET1928L be powered from an

IEC 61000-4-11

for 5 cycles

for 5 cycles

uninterruptible power supply or a battery.

 

70% UT

70% UT

 

 

(30% dip in UT)

(30% dip in UT)

 

 

for 25 cycles

for 25 cycles

 

 

<5% UT

<5% UT

 

 

(>95% dip in UT)

(>95% dip in UT)

 

 

for 25 cycles

for 25 cycles

 

 

 

 

 

Power frequency

3 A/m

3A/m

Power frequency magnetic fields

(50/60 Hz)

 

 

should be at levels characteristic of a

magnetic field

 

 

typical location in a typical commerical

 

 

 

or hospital environment.

IEC 61000-4-8

 

 

 

 

 

 

 

NOTE UT is the a.c. mains voltage prior to application of the test level.

 