·You can repeat the last two steps as often as necessary until you have selected all interference components (maximum 3) you need.

MLT25/CH2/R2

95.00 ppm

-- Channels --

MLT25/CH2

MLT25/CH3

MLT25/CH5

-

-

-

Measure

 

 

Back...

 

F1

F2

F3

F4

F5

4)Give the interference component into the sample gas channel and wait for a stable signal.

5)Go to the line "Calculate factor for selected interference component !" with the -key or the -key and then start the calculation with the 8 -key or the -key. If you want to see the results change to the line "Interference factors..." and press the 8 -key to go to the corresponding submenu:

 

TAG

95.00 ppm

 

 

 

 

1. Interference component factor:

1.000

 

2. Interference component factor:

1.000

 

 

Measure

 

Back...

F1

F3

F4

In this submenu you can also set up each interference component factor manual.

6)Select "Enabled" in the line "Compensation is" of the menu "Cross Interference Compensation" to start the cross interference compensation.

Function line "Remove selected component !"

If you press the 8 -key or the -key in this line you will delete the interference factor of the selected interference component !

5 - 32

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98