·You can repeat the last two steps as often as necessary until you have selected all interference components (maximum 3) you need.
MLT25/CH2/R2 | 95.00 ppm |
MLT25/CH2
MLT25/CH3
MLT25/CH5
-
-
-
Measure |
|
| Back... |
|
F1 | F2 | F3 | F4 | F5 |
4)Give the interference component into the sample gas channel and wait for a stable signal.
5)Go to the line "Calculate factor for selected interference component !" with the −
| TAG | 95.00 ppm |
|
|
|
| 1. Interference component factor: | 1.000 |
| 2. Interference component factor: | 1.000 |
|
|
Measure |
| Back... |
F1 | F3 | F4 |
In this submenu you can also set up each interference component factor manual.
6)Select "Enabled" in the line "Compensation is" of the menu "Cross Interference Compensation" to start the cross interference compensation.
Function line "Remove selected component !"
If you press the 8
5 - 32 | NGA 2000 |