5.1 Analyzer Module Setup

5.1.11 Concentration Peak Measurement

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

 

 

 

 

 

 

Peak measurement

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

-- Concentration Peak Measurement --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset minimum !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset maximum !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reset both !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Minimum detection:

 

 

 

 

 

 

 

 

Always

 

 

 

 

 

 

 

 

Maximum detection:

 

 

 

 

 

 

 

 

Always

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Difference (Max-Min):

 

 

 

 

 

 

 

2.500

 

 

 

 

 

 

 

 

Last minimum:

 

 

 

 

 

 

 

93.400

 

 

 

 

 

 

 

 

Last maximum:

 

 

 

 

 

 

 

95.900

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

 

Channel

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

 

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Concentration Peak Measurement" you can determine the conditions for the extreme values of a channel. If you want to change to the other channels of an MLT analyzer or analyzer module, you have to press the F3 key.

Set up parameters:

¨Select any line of variables or any function line with the -key or the -key.

¨Select the variable or start the function with the 8 -key or the -key.

¨If asked after function start: Confirm with the F2 key (Yes) or

cancel and go back to the menu page with the F4 key (Back...) or the -key.

¨Variable changing: Select the parameter with the -key or the -key.

¨Confirm the new parameter with the 8 -key or

cancel and go back to the last adjustment with the F2 key.

Function lines "Reset minimum !" or "Reset maximum !"

Starting these functions will reset the last minimum or maximum value of concentration. This is necessary to start a new minimum or maximum determination.

Function line "Reset both !"

Starting this function will reset the last minimum and the last maximum value of concentration simultaneously.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 55