5.1Analyzer Module Setup

5.1.14 Pressure Compensation

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

 

 

Pressure compensation

 

 

 

 

TAG

 

 

 

95.00 ppm

 

-- Pressure Compensation --

 

Compensation:

 

Use manual pressure

Manual pressure:

 

 

1013.0 hPa

Unit:

 

 

 

hPa

Pressure measurement is:

 

Valid

Reference pressure:

 

 

1013.0 hPa

Pressure:

 

 

 

1013.0 hPa

Measure

 

Channel

Back...

 

F1

F2

F3

F4

F5

In the menu "Pressure Compensation" you can set up the parameters of pressure for the current channel of an MLT analyzer or analyzer module. With the F3 key you can change to the other channels.

Set up parameters:

¨Select any line of variables with the -key or the -key.

¨Select the variable with the 8 -key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key

resp. select the whole parameter with the -key or the -key.

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.

Line of variables "Compensation":

¨Use manual pressure:

The pressure compensation of the current channel is based on the value of the line "Manual pressure".

¨Use sensor value:

The pressure compensation of the current channel is based on the value measured by a pressure sensor. The corresponding value will be displayed in the line "Reference pressure". This setup is not available, if no sensor is installed. Then, you will find the standard pressure of 1013.0 hPa resp. 14.7 psig in that line.

¨Disabled:

No pressure compensation will be done.