5.1Analyzer Module Setup

5.1.15 Flow Measurement

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

Flow measurement

TAG

95.00 ppm

-- Flow Measurement --

Unit:ml/min

Flow measurement is:

 

 

Valid

Flow:

 

 

 

23.0 ml/min

Measure

 

 

Back...

 

F1

F2

F3

F4

F5

In the menu "Flow Measurement" you can set up the flow unit for the current channel of an MLT analyzer or analyzer module. In the last line of this menu you will find the actual flow of the channel selected.

Flow unit setup:

¨Press the 8 -key or the -key to select the variable.

¨Select the unit with the -key or the -key. Options: ml/min, l/min

¨Confirm your adjustment with the 8 -key or cancel and go back to the last one with the F2 key.

Note:

If no flow sensor is installed, you will find a corresponding message on the screen instead of the menu page shown above.