5.1Analyzer Module Setup
5.1.17 Load/Save Analyzer Module Configuration
Main Menu — Analyzer and
↓
Analyzer module setup
↓[2nd menu page via F5 (More...)]
Loading/saving configuration parameters
↓
TAG | 95.00 ppm |
Send configuration to serial interface !
Load configuration from serial interface !
Reset analyzer module...
- BE CAREFUL with this function -
Replace current configuration with factory settings !
Measure |
|
| Back... |
|
F1 | F2 | F3 | F4 | F5 |
In the menu "Load/Save Analyzer Module Configuration" you can start several functions to send or load configuration data of the MLT analyzer or analyzer module via the serial interface. These functions are only available, if a SIO with serial interface is installed in the MLT analyzer or analyzer module.
Attention with loading of data:
You will delete all data of the RAM!
Starting the functions:
¨Select the function line you want with the ↓
¨Press the 8
cancel and go back to the menu page with the F4 key (Back...) or the ←
Function line "Send configuration to serial interface !":
You will send the RAM data via the serial interface of the MLT analyzer or analyzer module to the EPROM programmer or to an external computer!
Function line "Load configuration from serial interface !":
You will load data from an external memory to the RAM of the MLT analyzer or analyzer module via the serial interface. So you will delete the current RAM data!
Function line "Replace current configuration with factory settings !":
You will delete the RAM data and load the factory settings from the