·Go to the line you want with the -key or the -key.

·Select the tag of the channel with the 8 -key or the -key:

The display will go back to the menu "Analog Outputs" automatically. The tag of the channel selected will appear in the line "Analyzer module".

3)Select the signal name:

·Change to the line "Signal name" with the -key

·Select the parameter with the 8 -key or the -key and adjust one of the following parameters with the -key or the -key:

Concentration

Average

Peak-Value

Temperature

Pressure

Flow

Range

Calculator-1

Calculator-2

Calculator-3

Calculator-4

·Confirm the parameter selected with the 8 -key.

4)Determination of the output signal values:

·You can determine the signal value for the 0% output and the 100% output in the lines "Signal value for 0% output" or "Signal value for 100% output". So you have the possibility to zoom out a certain part of the whole range.

·Example:

Range from 0 to 1000 ppm

0% value shall be 400 ppm, 100% value shall be 700 ppm

Analog output is normally:

0V

= 0 ppm

10V

= 1000 ppm

After changing the output signal:

0V

= 400 ppm

10V

= 700 ppm

·Change to the line "Signal value for 0% output" or "Signal value for 100% output" with the -key.

·Select the value with the 8 -key or the -key.

·Select any digit with the -key or the -key and adjust a new value with the -key or the -key.

·Confirm the new value with the 8 -key.

·Note:

If you change the measurement range, the adjustment done in this menu will dis- appear and go back to the standard values of the range! To change the output signal values permanently you have to change the adjustments in the menu "Begin and End of Ranges" (see 5.1.3 p. 5-27)!

·Attention:

The signal range at the analog output should not be less than the littlest range! Otherwise, the noise level may be very high at the analog output!

5 - 70

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98