·Go to the line you want with the −
·Select the tag of the channel with the 8
The display will go back to the menu "Analog Outputs" automatically. The tag of the channel selected will appear in the line "Analyzer module".
3)Select the signal name:
·Change to the line "Signal name" with the ↓
·Select the parameter with the 8
„Concentration
„Average
„
„Temperature
„Pressure
„Flow
„Range
„
„
„
„
·Confirm the parameter selected with the 8
4)Determination of the output signal values:
·You can determine the signal value for the 0% output and the 100% output in the lines "Signal value for 0% output" or "Signal value for 100% output". So you have the possibility to zoom out a certain part of the whole range.
·Example:
„Range from 0 to 1000 ppm
„0% value shall be 400 ppm, 100% value shall be 700 ppm
„ Analog output is normally: | 0V | = 0 ppm | 10V | = 1000 ppm |
„ After changing the output signal: | 0V | = 400 ppm | 10V | = 700 ppm |
·Change to the line "Signal value for 0% output" or "Signal value for 100% output" with the ↓
·Select the value with the 8
·Select any digit with the ←
·Confirm the new value with the 8
·Note:
If you change the measurement range, the adjustment done in this menu will dis- appear and go back to the standard values of the range! To change the output signal values permanently you have to change the adjustments in the menu "Begin and End of Ranges" (see 5.1.3 p.
·Attention:
The signal range at the analog output should not be less than the littlest range! Otherwise, the noise level may be very high at the analog output!
5 - 70 | NGA 2000 |
90003482(2)