5.1 Analyzer Module Setup

5.1.18 Inputs and Outputs – Local SIO

Configuration of Relay Outputs:

Press the 8 -key or the -key in the line "Configuration of Relay Outputs..." of the menu "Local SIO Configuration Parameters" to change to the corresponding menu:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

-- Relay Outputs --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Relay 1 - output signal:

 

 

 

 

4

 

 

 

 

 

 

 

 

Relay 2 - output signal:

 

 

 

 

5

 

 

 

 

 

 

 

 

Relay 3 - output signal:

 

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Relay 1 - actual state:

 

 

 

 

 

 

On

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Relay 2 - actual state:

 

 

 

 

 

 

Off

 

 

 

 

 

 

 

 

Relay 3 - actual state:

 

 

 

 

 

 

Off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

 

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Relay Outputs" you can attach signals to the relay outputs of the local SIO board, e.g the relation between the zero valve and a certain channel. You have to put a number code (= variable of a signal) in the corresponding line for the relay:

See the table "Signal Codes 1 - 359" on the following pages: p. 5-75to 5-77 !

There are three relays on the local SIO board. With a jumper you can determine the relay contact status: NO or NC. Informations about the whole specification of the SIO board you can find in another manual.

Note:

In this menu you only can determine the relay output configuration for channels of the analyzer module, not for channels of auxiliary modules.

Set up parameters:

¨Select any line of variables with the -key or the -key.

¨Select the variable with the 8 -key or the -key.

¨Select the code number for the signal with the -key or the -key.

+ Options: Code 1 to 359 according to the list at page 5-75to 5-77 !

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.

Lines of variables "Relay 1, 2, 3 - actual state":

These lines display the current gear status of each relay. The lines are not available.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000