Function Codes 1 - 599

Available for:

DIO inputs (see p. 5-74)

The following functions occur only if the signal at the input changes from low to high level:

ID

 

 

 

General functions

1

Set all channels into STANDBY state (Stop procedures)

2 - 99

Reserved

 

 

 

 

 

 

 

 

 

ID

ID

ID

ID

ID

Channel dependent

CH1

CH2

CH3

CH4

CH5

functions

100

200

300

400

500

Start Zero calibration

101

201

301

401

501

Start Span calibration

102

202

302

402

502

Start Zero + Span calibration

103

203

303

403

503

Open sample gas valve

104

204

304

404

504

Open zero gas valve

105

205

305

405

505

Open purge gas valve

106

206

306

406

506

Open test gas valve

107

207

307

407

507

Open linearizer gas valve

108

208

308

408

508

Open span gas valve (of current range)

109

209

309

409

509

Open span gas valve of range-1

110

210

310

410

510

Open span gas valve of range-2

111

211

311

411

511

Open span gas valve of range-3

112

212

312

412

512

Open span gas valve of range-4

113

213

313

413

513

Close all valves (STANDBY)

114

214

314

414

514

Set range-1

115

215

315

415

515

Set range-2

116

216

316

416

516

Set range-3

117

217

317

417

517

Set range-4

118

218

318

418

518

Reserved

...

...

...

...

...

...

149

249

349

449

549

Reserved

The following functions are active if the signal at the input is set to high level:

ID

ID

ID

ID

ID

Channel dependent

CH1

CH2

CH3

CH4

CH5

functions

150

250

350

450

550

AK key-switch: Remote

151

251

351

451

551

Namur signal: Function Control

152

252

352

452

552

Reserved

...

...

...

...

...

...

199

299

399

499

599

Reserved

5 - 78

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98