5.1 Analyzer Module Setup

5.1.19 Delay and Average

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

 

 

 

 

 

 

Delay and average

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

 

-- Delay and Average --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Concentration output delay time:

 

 

 

 

1.000 s

 

 

 

 

 

 

 

 

Delay function is:

 

 

 

 

 

 

Disabled

 

 

 

 

 

 

 

 

Averaging time:

 

 

 

 

 

 

2 min

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Set average to 0.0 !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Concentration average:

 

 

 

 

 

93.87 ppm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

Channel

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Delay and Average" you can set up the

¨analog output delay time of the measurement signal and

¨averaging time to calculate the mean value of concentration

for one channel of an MLT analyzer or analyzer module. To set up the parameters of the other channels, you have to change with the F3 key.

Set up parameters:

¨Select any line of variables or the function line with the -key or the -key.

¨Select the variable or start the function with the 8 -key or the -key.

¨If asked after function start: Confirm with the F2 key (Yes) or

cancel and go back to the menu page with the F4 key (Back...) or the -key.

¨Variable changing:

Select any digit with the -key or the -key and adjust a new value with the -key or the -key

resp. select the whole parameter with the -key or the -key.

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 79