5.1Analyzer Module Setup

5.1.21 AK-Protocol Communication

Main Menu — Analyzer and I/O-module expert configuration

 

 

 

 

Analyzer module setup

 

 

[3rd menu page via 2 times F5 (More...)]

 

 

AK-protocol communication

 

 

 

 

 

 

 

 

 

 

 

 

TAG

95.00 ppm

 

 

-- AK-Communication --

 

 

 

 

 

Remote control:

Enabled

 

 

 

 

 

 

 

Remote control set to "FullAK" means:

-external range control is disabled

-automatic range switch level calculation is disabled

-t90 time does not change with ranges

-all valves closed if Operation status = Ready

- calibrations include SNGA/SEGA procedures

Measure

 

Channel

Back...

 

F1

F2

F3

F4

F5

In the menu "AK-Communication" you can set up the parameters for the remote control of the current channel via the serial interface. With the F3 key you can change to the other channels of an MLT analyzer or analyzer module.

Set up parameters:

¨Press the 8 -key or the -key to select the parameter.

¨Select the parameter you want with the -key or the -key.

Options:

·Disabled: No data transfer by remote control is possible.

·Full AK: The data transfer is only possible for commands of the AK standard.

·Enabled: The data transfer is possible for the commands of the AK standard and further commands like service commands.

¨Confirm the parameter selected with the 8 -key or cancel and go back to the last setup with the F2 key.

Notes:

¨Further informations about the "AK-communication protocol" you will find in an own manual.

¨To setup the data transfer parameters (e.g. baudrate) you have to change to the menu "Serial Interface" (see section 5.1.18 p. 5-72resp. 5.2.1 p. 89).

5 - 82

NGA 2000

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98