5.1Analyzer Module Setup

5.1.18 Inputs and Outputs – Local DIO

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

 

 

 

 

 

 

Inputs and outputs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIO-module(s)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TAG

 

 

 

 

 

 

 

 

 

95.00 ppm

 

 

 

 

 

 

 

 

 

 

 

-- Local DIO Boards Setup --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Number of selected input:

 

 

 

1

 

 

 

 

 

 

 

 

- Function code:

 

 

 

 

 

 

13

 

 

 

 

 

 

 

 

Number of selected output:

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

- Signal code:

 

 

 

 

 

 

111

 

 

 

 

 

 

 

 

Reset output fail !

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Board status:

 

 

 

 

 

 

 

Normal

 

 

 

 

 

 

 

 

Slot number of selected board:

 

 

 

6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Current status of selected input:

 

 

 

 

 

Low

 

 

 

 

 

 

 

 

Current status of selected output:

 

 

 

 

 

High

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

 

NEXT_DIO

 

Back...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

F2

F3

 

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Local DIO Boards Setup" you can set up the configuration of the local DIO modules of the analyzer module. Each DIO module is composed of 8 digital inputs and 24 digital outputs. You can attach a function to each input (e.g. opening of a valve) and a signal to each output (e.g zero valve) by a corresponding code.

See the tables on the following pages:

+Signal codes 1 -359: see p. 5-75to 5-77; Function codes: 1 - 599 see p. 5-78!

Set up parameters:

¨Select the line of variables or the function line with the -key or the -key.

¨Select the variable or start the function with the 8 -key or the -key.

¨If asked after function start: Confirm with the F2 key (Yes) or

cancel and go back to the menu page with the F4 key (Back...) or the -key.

¨Variable changing:

Select any digit with the -key or the -key and adjust a new value with the -key or the -key.

¨Confirm the new value with the 8 -key or

cancel and go back to the last value with the F2 key.