5.1 Analyzer Module Setup
5.1.18 Inputs and Outputs – Local SIO
5)Determination of the output current range:
·Press the 8
·Confirm the new adjustment with the 8
·Options: 0...20 mA (and automatically 0...10V) or 4...20 mA (2...10V).
6)Concentration scaling:
·Press the 8
·Confirm the new parameter with the 8
7)Fine adjustment of the analog output:
·If you press the F5 key (More...) in the menu "Analog Outputs", you can change to the submenu "Fine Adjustment", where you can do the fine adjustments of the analog output:
TAG |
|
|
| 95.00 ppm |
|
|
| ||
Output number: |
|
| 1 | |
Operation mode: |
|
| Normal | |
Fine adjustment for 0% output: |
| 4097 | ||
Fine adjustment for 100% output: |
| 799 | ||
Measure | Updates | Back... |
| |
F1 | F2 | F3 | F4 | F5 |
·Press the 8
„Normal: The absolute measurement signal will be sent to the analog output.
„Zero: Coordination between the display and the analog output for 0V with fine adjustment 0%.
„Full Scale: Coordination between the display and the analog output for 10V with fine adjustment 100%.
·Confirm the new parameter with the 8
·Press the 8
·Adjust the value you want with the −
The last three lines of variables in the menu "Analog Outputs" only display some values of the analog output! They are not available.