5.1 Analyzer Module Setup

5.1.13 Gasflow Setup

Main Menu — Analyzer and I/O-module expert configuration

Analyzer module setup

[2nd menu page via F5 (More...)]

Gasflow setup

TAG

 

 

 

95.00 ppm

 

 

-- Gasflow Setup --

 

Gasflow through analyzer cell is:

 

Parallel

Valve position at "Basic-Status":

 

Samplegas

Valve position:

 

 

Samplegas

Operation status:

 

 

Ready

Measure

 

Channel

Back...

 

F1

F2

F3

F4

F5

In the menu "Gasflow Setup" you can set up the modus of gas flow through the analyzer cell:

¨Parallel or

¨Serial

and the valve position at "Basic-Status":

¨Samplegas or

¨All closed

Notes:

¨The adjustment for the gas flow through the analyzer cell depends on the real combination of the analyzer cells.

¨The setup is normally correct by factory setting.

You have only to change it, if you modify your configuration of cells.

¨The correctness of this adjustment is very important for the time controlled calibration with valve gear (see 5.1.1 p. 5-12)

Set up parameters:

¨Change to the line you want with the -key or the -key.

¨Press the 8 -key or the -key to select the variable.

¨Select the parameter you want with the -key or the -key.

¨Confirm your adjustment with the 8 -key or cancel and go back to the last one with the F2 key.

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98

NGA 2000

5 - 59