5.1 Analyzer Module Setup
5.1.13 Gasflow Setup
Main Menu — Analyzer and
↓
Analyzer module setup
↓ [2nd menu page via F5 (More...)]
Gasflow setup
↓
TAG |
|
|
| 95.00 ppm |
|
|
| ||
Gasflow through analyzer cell is: |
| Parallel | ||
Valve position at |
| Samplegas | ||
Valve position: |
|
| Samplegas | |
Operation status: |
|
| Ready | |
Measure |
| Channel | Back... |
|
F1 | F2 | F3 | F4 | F5 |
In the menu "Gasflow Setup" you can set up the modus of gas flow through the analyzer cell:
¨Parallel or
¨Serial
and the valve position at
¨Samplegas or
¨All closed
Notes:
¨The adjustment for the gas flow through the analyzer cell depends on the real combination of the analyzer cells.
¨The setup is normally correct by factory setting.
You have only to change it, if you modify your configuration of cells.
¨The correctness of this adjustment is very important for the time controlled calibration with valve gear (see 5.1.1 p.
Set up parameters:
¨Change to the line you want with the −
¨Press the 8
¨Select the parameter you want with the −
¨Confirm your adjustment with the 8