6.5 System Reset

Main Menu — System configuration and diagnostics

System Reset

TAG

37.50 ppm

-- System Reset --

Are you sure ???

Re-initializing will destroy all the binds !!!

System reset !

Re-initialize network !

Measure

 

 

Back...

 

F1

F2

F3

F4

F5

You can start the following functions in the menu "System Reset":

1)System Reset:

·The MLT analyzer or the platform will change to the initializing mode automatically, if you press the 8 -key or the -key in the line "System reset !":

This reboot is analogous to the switching on of the control module.

2)Re-initialize network:

·Change to the line "Re-initialize network!" with the the -key or the -key.

·Press the 8 -key or the -key to start this function.

Caution:

¨With the function "Re-initialize network !", you will delete all bindings between analyzer modules and I/O modules.

¨All configuration data of the SIO and DIO modules will disappear!

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98