The result of calibration for included AMs is displayed in an own menu page. This can be achieved by softkey „Result...“ from different menu pages of SYSCAL.
- Calibration Results - |
|
|
|
|
|
Choose analyzer module... |
| SCMODULE |
CLD | ||
Enabled for | No | SCCONTROL |
_______________________________________________ | SCLRESULT | |
Result of last run: | OK | |
Successfully calibrated ranges: | 1+2+4 | SCVALIDITY |
NO | SCRESULT1 | |
YES | SCRESULT2 | |
|
|
|
Remember to run the calibration procedures as a test without actually to calibrate assert parameter "Calibration Procedures in
It is also possible to switch for test purposes onto a special gas of a specific module. This possibility is located in menu "Test procedures".
System configuration and diagnostics...
↓
System Calibration...
↓
Calibration/Test Procedures...
↓
Test procedures...
↓
| |
|
|
Gastest of specific module | |
Timeout for Gastest (0 = no timeout) | 180 s |
Choose specific analyzer module... |
|
_______________________________________________
Procedure time:10 s
SCTESTGAS SCTIMEOUT
SCCALTIME1 SCSTEPMOD
90003482(2) | NGA 2000 | Supplement - 15 |