NGA-2000 System Calibration

The result of calibration for included AMs is displayed in an own menu page. This can be achieved by softkey „Result...“ from different menu pages of SYSCAL.

- Calibration Results -

 

 

 

 

 

Choose analyzer module...

 

SCMODULE

Analyzer-module:

CLD

Enabled for system-cal:

No

SCCONTROL

_______________________________________________

SCLRESULT

Result of last run:

OK

Successfully calibrated ranges:

1+2+4

SCVALIDITY

Zero-Cal fail of any analyzer module:

NO

SCRESULT1

Span-Cal fail of any analyzer module:

YES

SCRESULT2

 

 

 

Remember to run the calibration procedures as a test without actually to calibrate assert parameter "Calibration Procedures in Test-Mode" to <Yes>.

It is also possible to switch for test purposes onto a special gas of a specific module. This possibility is located in menu "Test procedures".

System configuration and diagnostics...

System Calibration...

Calibration/Test Procedures...

Test procedures...

-- Test Procedures --

 

 

 

Gastest of specific module

SAMPLE-Gas

Timeout for Gastest (0 = no timeout)

180 s

Choose specific analyzer module...

 

_______________________________________________

Procedure time:10 s

Analyzer-module:MLT/CH3

SCTESTGAS SCTIMEOUT

SCCALTIME1 SCSTEPMOD

90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98