6.3.2.4 Diagnostic Menus

Temperature Control

Main Menu — Technical level configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diagnostic menus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module diagnostics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temperature control parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Temperature control

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Converter set point:

 

 

 

 

 

300.0 C

 

 

 

 

 

 

 

 

 

Converter P gain:

 

 

 

 

 

0.015

 

 

 

 

 

 

 

 

 

Converter I gain:

 

 

 

 

 

0.000125

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Converter bias:

 

 

 

 

 

0.6

 

 

 

 

 

 

 

 

 

Converter temperature:

 

 

 

304

 

 

 

 

 

 

 

 

 

Detector set point:

 

 

 

 

 

51.5

 

 

 

 

 

 

 

 

 

Detector P gain:

 

 

 

 

 

0.100

 

 

 

 

 

 

 

 

 

Detector I gain:

 

 

 

 

 

0.00170

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector bias:

 

 

 

 

 

0.4 C

 

 

 

 

 

 

 

 

Detector temperature:

 

 

 

 

 

25.8 C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

MORE

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

F3

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "Temperature control" you will find the parameters of the converter's and detector's PID controller. Besides, you can set up the converter set point in this menu. So, you can optimize the effectiveness of the converter.

If there are more than ten parameters available, you can change to another menu page with the F3 key.

Set up parameters:

¨Press the ENTER key or the -key in the line "Converter set point" to select the parameter.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key

Options: 0 to 420 °C; Standard: 300 °C to 380 °C.

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98

NGA 2000

6 - 35