6.3.2.5 Diagnostic Menus

Flow control parameters

Main Menu — Technical level configuration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Diagnostic menus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyzer module diagnostics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow control parameters

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLD

 

 

 

 

 

 

 

 

95.0 ppm

 

 

 

 

 

 

 

 

 

 

 

 

NO/NOx flow balance

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOx correction factor:

 

 

 

1.0000

 

 

 

 

 

 

 

 

 

Calculate factor using pressure ratio...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Calculate factor using span gas response ratio...

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HOME

 

ESCAPE

 

 

 

 

 

INFO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

F1

 

F2

 

F3

F4

F5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

In the menu "NO/NOx flow balance" you can set up or calculate an NOx to NO flow correction factor.

Since the sample flows through different paths in the NO and NOx modes of analysis, there are differences in the flows entering the reaction chamber in the two modes. So the CLD may produce wrong results of analysis. To compensate the different flow rate you can set up a correction factor manually or calculate it in two different methods.

Note:

¨This menu page will only be available in the software revision 2.3! The correction function will only be valid for the CLD hardware "CLD 2" without flow balance valve!

¨The NO/NOx compensation of the former models will have to be done with the flow balance valve! In that case, you will not find the menu page shown above.

Set up parameters:

¨Select the line of variables or the menu line with the -key or the -key.

¨Select the variable or change to the submenu with the ENTER key or the -key.

¨Select any digit with the -key or the -key and adjust a new value with the -key or the -key.

¨Confirm the new value with the ENTER key or cancel and go back to the last value with the F2 key.

6 - 36

NGA 2000

90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98