System Calibration

2 Valves for system calibration

Before using the system calibration we have to put together the required valve pool. There are three types of I/O-modules supporting this:

DIO

- 24 digital outputs / 8 digital inputs (max. 4 modules per CM)

SIO

- 3 digital outputs (max. 1 module per CM)

[CVU - 4 digital outputs (max. 4 modules per CM)]

Software supports up to 16 system valves.

2.1 Assigning an output-port to a system valve

Assigning of an output can be accomplished by using the menus for selected output module (DIO, SIO or CVU).

There we have to select the NGA Control Module as the Source Module. Control module then provides the signals for system valve V1...V16.

For example the DIO:

Expert controls and set up...

Auxiliary module controls...

Set-up DIO module(s)...

- DIO-MODULE OUTPUTS-

Inputs...

 

Outputnumber:

1

Choose module...

 

Choose signal...

 

Invert output:

DISABLED

Module status:

NORMAL

Slot ID:

1

Signal name:

SYS:VALVE-1

Signal level:

OFF

Signal comes from:

NGA Control Module

DIOOUTNUMC

DIOOUTINVC DIOMODSTAC DIOSLOTIDC DIOOUTSIGC DIOOUTSTATC DIOOUTSRCC

There we have to

select "Outputnumber"

then choose „NGA Control Module“ as module

then choose the wanted valve „SYS: VALVE-x“ as signal

Supplement - 4