System Calibration
2 Valves for system calibration
Before using the system calibration we have to put together the required valve pool. There are three types of
∙ | DIO | - 24 digital outputs / 8 digital inputs (max. 4 modules per CM) |
∙ | SIO | - 3 digital outputs (max. 1 module per CM) |
∙[CVU - 4 digital outputs (max. 4 modules per CM)]
Software supports up to 16 system valves.
2.1 Assigning an
Assigning of an output can be accomplished by using the menus for selected output module (DIO, SIO or CVU).
There we have to select the NGA Control Module as the Source Module. Control module then provides the signals for system valve V1...V16.
For example the DIO:
Expert controls and set up...
↓
Auxiliary module controls...
↓
↓
-
Inputs... |
|
Outputnumber: | 1 |
Choose module... |
|
Choose signal... |
|
Invert output: | DISABLED |
Module status: | NORMAL |
Slot ID: | 1 |
Signal name: | |
Signal level: | OFF |
Signal comes from: | NGA Control Module |
DIOOUTNUMC
DIOOUTINVC DIOMODSTAC DIOSLOTIDC DIOOUTSIGC DIOOUTSTATC DIOOUTSRCC
There we have to
∙select "Outputnumber"
∙then choose „NGA Control Module“ as module
∙then choose the wanted valve „SYS:
Supplement - 4 | NGA 2000 |