System Calibration

It is also possible to configure the DIO-outputs via LON-Variables (i.e. SLTA-adapter or AK-protocol).

Therefore is the LON-variable "DIOOUT_ENTRYSIG" which corresponds to DIOOUTSIGC and evaluates enum-values of ST1NAME (enum-values 0..19) or ST2NAME (enum-values 20..39).

For digital output we have to setup variables in following order.

1.

DIOSLOTIDC

 

2.

DIOOUTNUMC

 

3.

DIOOUTSRCC

 

4.

DIOOUT_ENTRYSIG SYS: VALVE-1

= 20 (enum-value in ST2NAME)

 

SYS: VALVE-2

= 21

 

.

.

 

.

.

 

.

.

 

SYS:VALVE-16

= 35

For each needed system valve we have to repeat these appropriate selections.

It is also possible to distribute the system valves onto different output modules.

90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98

NGA 2000