System Calibration

3.3 Set-up general parameters

One general parameter is "Calibration Procedures in Test-Mode". It is located in the menu "Calibration/Test procedures".

Expert controls and set up...

System Calibration...

Calibration/Test procedures...

Calibration/Test Procedures

 

 

Start Zeroing all ranges !

 

 

Start Zeroing and Spanning all ranges !

 

 

Start calibration program !

 

 

Cancel calibration !

 

SCTESTMOD

Calibration Procedures in Test-Mode:

No

 

Test Procedures...

 

SCCALTYPE2

Calibration type:

Zero-Cal

SCPROGSTEP

Program step:

1

SCCALTIME1

Calibration time:

16 s

SCCALTIME2

Previous calibration time:

57 s

 

 

 

 

 

 

RESULTS

 

 

 

With this parameter we can run all the calibration procedures either in the defined mode or in a test mode. The test mode means that the valve switching and waiting for purge times is done in the same manner like in the normal calibration procedure.

The only difference is that the single calibrations of the modules and the appropriate times the modules would need for the calibration are not done.

90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98

NGA 2000

Supplement - 11