System Calibration
3.3
One general parameter is "Calibration Procedures in
Expert controls and set up...
↓
System Calibration...
↓
Calibration/Test procedures...
↓
Calibration/Test Procedures |
|
| |||
Start Zeroing all ranges ! |
|
| |||
Start Zeroing and Spanning all ranges ! |
|
| |||
Start calibration program ! |
|
| |||
Cancel calibration ! |
| SCTESTMOD | |||
Calibration Procedures in | No | ||||
| |||||
Test Procedures... |
| SCCALTYPE2 | |||
Calibration type: | |||||
SCPROGSTEP | |||||
Program step: | 1 | ||||
SCCALTIME1 | |||||
Calibration time: | 16 s | ||||
SCCALTIME2 | |||||
Previous calibration time: | 57 s | ||||
| |||||
|
|
|
| ||
| RESULTS |
|
|
|
With this parameter we can run all the calibration procedures either in the defined mode or in a test mode. The test mode means that the valve switching and waiting for purge times is done in the same manner like in the normal calibration procedure.
The only difference is that the single calibrations of the modules and the appropriate times the modules would need for the calibration are not done.
90003749(1) | NGA 2000 | Supplement - 11 |