System Calibration

3.2 Programming Calibration Sequences

Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to run the system calibration in an user definable order of up to 40 steps.

The menu to set-up this feature looks as follows.

Expert controls and set up...

System Calibration...

Calibration sequence programming...

Calibration Sequence Programming

Step #:

1

Choose specific analyzer module...

 

Select all analyzer modules !

 

Calibration procedure type:

Zero-Cal

Analyzer module:

MLT/CH3

Prg1-10 Prg11-20 Prg21-30 Prg31-40

SCSTEP

SCCALTYPE1

SCSTEPMOD

For programming the sequence you must

1.select „Step #“

2.select „Calibration procedure type“

3.select module

Repeat this order for all program steps.

The programmable „Calibration procedure types“ are:

1.

NOP

no operation (for deleting a step in an existing program)

2.

ZERO-CAL

do a zero calibration

3.

SPAN-CAL

do a span-calibration for all available ranges

4.

ZERO/SPAN-CAL

do a zero-cal then a span-cal for all available ranges

5.

SPAN1-CAL

do a span calibration only for range #1

6.

SPAN2-CAL

do a span calibration only for range #2

7.

SPAN3-CAL

do a span calibration only for range #3

8.

SPAN4-CAL

do a span calibration only for range #4

9.

END-OF-PROG

end of sequence

90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98

NGA 2000

Supplement - 9