System Calibration
3.2 Programming Calibration Sequences
Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to run the system calibration in an user definable order of up to 40 steps.
The menu to
Expert controls and set up...
↓
System Calibration...
↓
Calibration sequence programming...
↓
Calibration Sequence Programming
Step #: | 1 |
Choose specific analyzer module... |
|
Select all analyzer modules ! |
|
Calibration procedure type: |
Analyzer module: | MLT/CH3 |
SCSTEP
SCCALTYPE1
SCSTEPMOD
For programming the sequence you must
1.select „Step #“
2.select „Calibration procedure type“
3.select module
Repeat this order for all program steps.
The programmable „Calibration procedure types“ are:
1. | NOP | no operation (for deleting a step in an existing program) |
2. | do a zero calibration | |
3. | do a | |
4. | do a | |
5. | do a span calibration only for range #1 | |
6. | do a span calibration only for range #2 | |
7. | do a span calibration only for range #3 | |
8. | do a span calibration only for range #4 | |
9. | end of sequence |