System Calibration
3.4 Controlling System Calibration
After
system
In this mode a
system
In this mode will be performed a
user defined sequence program: In this mode the user is responsible for the
| optimization. See setup of this mode. |
gas test: | It is also possible to switch for test purposes onto a |
| specific gas of a specific module. A calibration is not |
| done. |
Any mode can be started by the following instances:
∙Manually by operator interaction
∙Triggered by programmable input (DIO)
∙
∙Programmed
∙
It is also possible to cancel a running system calibration. This can be done by the following instances:
∙Manually by operator interaction
∙Triggered by programmable input (DIO)
∙
∙
There is no priority by which way SYSCAL can be started. If SYSCAL was started it can not be restarted by a further instance. Only after canceling it can be restarted.
90003749(1) | NGA 2000 | Supplement - 13 |