System Calibration

The result of calibration for included AMs is displayed in an own menu page. This can be achieved by softkey „RESULTS“ from different menu pages of SYSCAL.

Calibration Results

 

 

 

 

 

Change analyzer module...

 

SCMODULE

Analyzer module:

CLD

Enabled for system-cal:

No

SCCONTROL

Successful zero+span calibrated ranges:

1+2+4

SCVALIDITY

Zero-Calibration failed on any analyzer module:

No

SCRESULT1

Span-Calibration failed on any analyzer module:

Yes

SCRESULT2

Remember to run the calibration procedures as a test without actually to calibrate assert parameter "Calibration Procedures in Test-Mode" to <Yes>.

It is also possible to switch for test purposes onto a specific gas of a specific module. This possibility is located in menu "Test procedures".

Expert controls and set up...

System Calibration...

Calibration/Test Procedures...

Test procedures...

Test Procedures

 

 

 

Gastest of specific module

Sample-Gas

Timeout for gastest (0 = no timeout)

180 s

Choose specific analyzer module...

Analyzer-module:MLT/CH3

SCTESTGAS SCTIMEOUT

SCSTEPMOD

90003749(1) [NGA-e (CLD-Software 2.3/MLT-Software 3.1.X)] 04/98

NGA 2000

Supplement - 15