System Calibration
The result of calibration for included AMs is displayed in an own menu page. This can be achieved by softkey „RESULTS“ from different menu pages of SYSCAL.
Calibration Results |
|
|
|
|
|
Change analyzer module... |
| SCMODULE |
Analyzer module: | CLD | |
Enabled for | No | SCCONTROL |
Successful zero+span calibrated ranges: | 1+2+4 | SCVALIDITY |
No | SCRESULT1 | |
Yes | SCRESULT2 |
Remember to run the calibration procedures as a test without actually to calibrate assert parameter "Calibration Procedures in
It is also possible to switch for test purposes onto a specific gas of a specific module. This possibility is located in menu "Test procedures".
Expert controls and set up...
↓
System Calibration...
↓
Calibration/Test Procedures...
↓
Test procedures...
↓
Test Procedures |
|
|
|
Gastest of specific module | |
Timeout for gastest (0 = no timeout) | 180 s |
Choose specific analyzer module...
SCTESTGAS SCTIMEOUT
SCSTEPMOD
90003749(1) | NGA 2000 | Supplement - 15 |