Supplement
ASVC Kn S615 b | Reading of the current data of a DIO board | ||
Description: | The current status of the named DIO board will be read. | ||
Syntax: | b: Board number (1, 2, 3 ,4; not | ||
| for b = 0 the data of all available DIO's will be sent | ||
Response: | ASVC 0 IIIIIIII OOOOOOOO OOOOOOOO OOOOOOOO RLLL | ||
| I1...8 Status of the digital input pins 1...8 | ||
| O1...24 Status of the digital output pins 1...24 | ||
| R | Retrigger error |
|
| L1...3 | Overload group 1...3 |
|
|
| ||
SSVC Kx S617 n s [n s] | Setup of the external switch variable DIGEXTSWITCH | ||
Description: | The outputs of a DIO board can be assigned to the single bits of the variable DIGEXTSWITCH. | ||
| For the platform the lowest 8 bits of this variable are available. With the command "SSVC" | ||
| each switch can be set. This on the other hand may cause a direct switch of an assigned | ||
| digital output. |
| |
Syntax: | n: | Switch number |
|
| s = | Status (0 = OFF; 1 = ON) |
x:If the channel number will be K0, the variable DIGEXTSWITCH will be called of the module containing the local SIO
(platform SIO → control module, MLT SIO → analyzer module). The parameters "n s" can be repeated up to seven times for the platform.
Examples: | SSVC K0 S617 1 1 2 0 3 1 | Set external switch 1 to HIGH |
|
| Set external switch 2 to LOW |
|
| Set external switch 3 to HIGH |
90003752(1) | AK | Supplement - 5 |