ASVC Kn S640 a

Output of values of the MLT concentration formula

Concentration formula:

Conce = FacP * FacT * FacSpan * Lin{(RawAvg - OffP - OffT - OffX) * RGain * Gain}

Meaning of the variables:

RawAvg:

Raw value average (including t90 time)

OffP:

Offset correction of physics

 

OffT:

Offset correction of temperature

 

OffX:

Offset correction of interferences from other channels

RGain:

Analog pre amplifying factor (BIS)

Gain:

Factor standardizing to the desired setpoint value (span)

Lin:

Linearization procedure

 

FacSpan:

Span factor standardizing the product "FacP*FacT*FacSpan" to 1.0 during the span cal.

 

FacT:

Temperature correction in the sensitivity

 

FacP:

Pressure correction in the sensitivity

 

Conce:

Concentration value

Syntax:

ASVC Kn S640 a

 

Response:

ASVC 0 <value1> <value2> ...

 

 

a = 0:

value1 = RawCount

= uncorrected signal

 

 

value2 = RawAvg

 

 

 

value3 = OffT

 

 

 

value4 = OffX

 

 

 

value5 = OffP

 

 

 

value6 = Gain

 

 

 

value7 = RGain

 

 

a = 1:

value1 = LinInput = (RawAvg - OffP - OffT - OffX) * RGain * Gain

 

 

value2 = LinOutput = Lin{(RawAvg - OffP - OffT - OffX) * RGain * Gain}

 

 

value3 = FacSpan

 

 

 

value4 = FacT

 

 

 

value5 = FacP

 

 

 

value6 = Conce

 

 

a = 2:

value1 = Temperature for OffT

 

 

value2 = Temperature for FacT

 

 

value3 = Atmospheric pressure for FacP

Supplement - 8

AK